Background: No indication are available for transcatheter aortic valve implantation (TAVI) sizing in bicuspid aortic valve (BAV). Aim of the study is to develop and validate a Multi-Slice Computed Tomography (MSCT)-based algorithm for transcatheter heart valve (THV) sizing in patients with stenotic BAV under evaluation for TAVI.

Methods: A two steps method was applied: 1)evaluation of a cohort of 19 consecutive patients with type I BAV stenosis undergoing TAVI through pre and post-procedural MSCT, and development of an algorithm for THV sizing; 2)validation of the algorithm on a new cohort of 21 patients.

Results: In the first cohort, a high correlation was found between the raphe-level area measured at pre-procedural MSCT and the smallest THV area measured at post-procedural MSCT (p < 0.001). Moreover, reduced THV expansion was observed among patients with higher calcium burden (p = 0.048). Then, a new algorithm for TAVI sizing in BAV was develop (CASPER: Calcium Algorithm Sizing for bicusPid Evaluation with Raphe). This algorithm is based on the reassessment of the perimeter/area derived annulus diameter, according to three main anatomical features: 1) the ratio between raphe length and annulus diameter; 2)calcium burden; 3)calcium distribution in relation to the raphe. The algorithm was then validated in a new cohort of 21 patients, achieving 100% of procedural success and excellent TAVI performance.

Conclusion: MSCT assessment of raphe length, calcium burden and its distribution is of crucial relevance in the pre-procedural evaluation of patients with BAV. These anatomical features can be combined in a new and simple algorithm for TAVI sizing.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcct.2020.01.007DOI Listing

Publication Analysis

Top Keywords

aortic valve
16
bicuspid aortic
8
transcatheter aortic
8
valve implantation
8
multi-slice computed
8
computed tomography
8
thv sizing
8
post-procedural msct
8
area measured
8
valve
5

Similar Publications

Background: Conduction disturbances are common after transcatheter aortic valve implantation (TAVI) and frequently require permanent pacemaker implantation (PPI). Data regarding its impact on mortality and morbidity are conflicting. This study aims to assess the impact of PPI before or within 30 days after TAVI on mortality and health-related Quality of Life (QoL) during the first year after TAVI.

View Article and Find Full Text PDF

A 70-year-old man developed intermittent fever with chills, severe anorexia, generalized weakness, and mild exertional difficulty in breathing following posterior chamber intraocular lens replacement surgery for a mature white cataract in the left eye. Laboratory tests revealed persistent negative blood cultures, normocytic and normochromic anemia, neutrophilia, and elevated inflammatory markers despite multiple courses of antibiotics. All other investigations conducted to identify the cause of prolonged fever, including transthoracic echocardiography, were negative.

View Article and Find Full Text PDF

Background: The initial outcomes of transcatheter aortic valve replacement in patients with left ventricular outflow tract calcification are poor. Furthermore, balloon-expandable transcatheter aortic valve replacement is associated with an increased risk of annular rupture, and self-expandable transcatheter aortic valve replacement is associated with worse post-operative residual paravalvular leakage grades. Therefore, developing an optimal method for transcatheter aortic valve replacement for patients with left ventricular outflow tract calcification is desirable.

View Article and Find Full Text PDF