Filgotinib is a selective JAK1 (Janus kinase) inhibitor, filed in Japan for the treatment of rheumatoid arthritis. In this paper, we report a validated liquid chromatography coupled with tandem mass spectrometry for the quantification of filgotinib in rat plasma using tofacitinib as an internal standard (IS) as per the Food and Drug Administration regulatory guidelines. Filgotinib and the IS were extracted from rat plasma using ethyl acetate as an extraction solvent and chromatographed using an isocratic mobile phase (0.2% formic acid:acetonitrile; 20:80, v/v) at a flow rate of 0.9 mL/min on a Gemini C column. Filgotinib and the IS were eluted at ~1.31 and 0.89 min, respectively. The MS/MS ion transitions monitored were m/z 426.3 → 291.3 and m/z 313.2 → 149.2 for filgotinib and the IS, respectively. The calibration range was 0.78-1924 ng/mL. No matrix effect and carryover were observed. Intra- and inter-day accuracies and precisions were within the acceptance range. Filgotinib was stable for three freeze-thaw cycles: on bench-top up to 6 h, in an autosampler up to 21 h, and at -80°C for 1 month. This novel method has been applied to a pharmacokinetic study in rats.