Purpose: To perform dosimetric validation of the plan adaptation and high-definition motion management (HDMM) system of Gamma Knife® Icon in various clinical scenarios.
Methods And Materials: We built an assembly for a pitch-adjustable anthropomorphic head phantom. We then used films to measure dosimetric and positional accuracy in 13 clinical scenarios, including movement near HDMM thresholds, multiple plan adaptations, frequent coughing, and initial setup error.
Results: The dose for the superiorly located 4-mm shot was decreased up to 7-13% near 2- to 3-mm HDMM thresholds in the chin-down position. Dosimetric deviation was within ±3.5% for initial pitch angles of up to 20°. Multiple treatment interruption and frequent coughing did not cause substantial dosimetric deviation (<2%).
Conclusion: Our results indicated that dosimetric accuracy of the Gamma Knife® Icon system is reliable even in extreme treatment conditions. However, the user should exercise caution for superiorly located small lesions with an HDMM threshold ≥2 mm or in the scenario of large initial setup error.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6774480 | PMC |
Phys Med Biol
January 2025
Joint Department of Physics, The Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, Downs Road, Sutton, London, Surrey, SM2 5PT, UNITED KINGDOM OF GREAT BRITAIN AND NORTHERN IRELAND.
This study aims to develop and evaluate a fast and robust deep learningbased auto-segmentation approach for organs at risk in MRI-guided radiotherapy of pancreatic cancer to overcome the problems of time-intensive manual contouring in online adaptive workflows. The research focuses on implementing novel data augmentation techniques to address the challenges posed by limited datasets. Approach: This study was conducted in two phases.
View Article and Find Full Text PDFRadiol Phys Technol
January 2025
Department of Radiotherapy, Sri Shankara Cancer Hospital and Research Centre, 1st Cross, Shankarapuram, Basavanagui, Bengaluru, 560004, Karnataka, India.
This study evaluates the dosimetric impact of arc simulation angular resolution in VMAT-based Single Isocentre Multiple Target (SIMT) SRS, focusing on their dependence on target size, isocentre distance, number of arcs, and arc type. A phantom study analysed angular resolution (0.5°, 1°, 2°) effects on dosimetric accuracy for PTVs of 0.
View Article and Find Full Text PDFClin Transl Radiat Oncol
January 2025
Department of Radiotherapy, Erasmus MC Cancer Institute, University Medical Center Rotterdam, Dr. Molewaterplein 40, 3015 GD Rotterdam, the Netherlands.
Front Oncol
December 2024
Department of Radiation Oncology, Shandong Cancer Hospital and Institute, Shandong First Medical University, and Shandong Academy of Medical Sciences, Jinan, Shandong, China.
Sci Rep
December 2024
Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, Hangzhou, 310000, China.
Recently, neoadjuvant short-course radiation therapy (SCRT) has emerged as a valid treatment option for patients with locally advanced rectal cancer (LARC). We assessed SCRT plans using volumetric-modulated arc therapy (VMAT) with Halcyon and Infinity medical linear accelerators (Linacs) and compared the plan quality and delivery efficiency across all cases. Thirty patients who underwent preoperative SCRT for LARC at the hospital were randomly selected.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!