Purpose: The aim of this research was to present a new approach for the potential characteristics of the anterior eye, based on numerical analysis of the raw data from multiple measurements on 1 eye captured with use of the Ocular Response Analyzer (ORA).

Methods: The measurements were carried out 6 times on every eye, one by one, on 29 patients using the ORA. Dependencies between values of 6 applanation pressures P and P from 1 eye and the new proposed time-dependent parameters sm and smm, obtained from the raw data, were used to determine new characteristic features of the anterior eye.

Results: Obtained values for multiple applanation pressures P and P and their average value P from all measurements on 1 eye show strong linear dependencies with the new proposed time-dependent parameter smm.

Conclusions: The new parameters based on slope coefficients a and a, their difference Δa, and respective y-intercept values b, b, and Δb show some characteristic stability for a multiply measured eye and can be used for its characterization. These parameters can be used for differentiation of the eyes with higher accuracy than those obtained from a single measurement. They are likely related to some biomechanical properties of the anterior eye.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6945948PMC
http://dx.doi.org/10.1159/000452124DOI Listing

Publication Analysis

Top Keywords

raw data
12
analysis raw
8
ocular response
8
response analyzer
8
anterior eye
8
measurements eye
8
applanation pressures
8
proposed time-dependent
8
eye
7
approach analysis
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!