Background: Testing of protective sensation and vibration perception are two of the most commonly used non-invasive methods of screening for diabetes-related peripheral neuropathy (DPN). However, there is limited research investigating the reliability of these tests in people with diabetes. The aim of this study was to determine the inter- and intra-rater reliability of methods used to test vibration perception and protective sensation in a community-based population of adults with type 2 diabetes.

Methods: Three podiatrists with varying clinical experience tested four- and 10-site, 10 g monofilament and vibration perception threshold (VPT). In a separate cohort, the reliability of a graduated tuning fork as well as two methods of conventional tuning fork (on/off method and dampening method) was undertaken by a new graduate podiatrist and podiatrist with one-year's clinical experience. The intra- (Cohen's К) and inter-rater (Cohen's or Fleiss' К) reliability of each test was determined.

Results: Fifty participants (66% male, 100% type 2, 32% with DPN) underwent monofilament and neurothesiometer testing with 44 returning for the retest. Twenty-four participants (63% male, 100% type 2, 4% with DPN) underwent tuning fork testing and returned for retest. All tests demonstrated acceptable inter-rater reliability ranging from moderate (10-site monofilament, К: 0.54, CI: 0.38-0.70,  = 0.02) to substantial (graduated tuning fork, К: 0.68, CI: 0.41-0.95,  < 0.01). The 10-site monofilament (К: 0.44-0.77) outperformed the 4-site test (К: 0.34-0.67) and the dampened tuning fork method (К: 0.41-0.49) showed lower intra-rater reliability compared to both conventional (К: 0.52-0.57) and graduated methods (К: 0.50-0.57).

Conclusion: We support the current recommendations of using more than one test to screen and monitor progression of DPN. Four- and 10-site 10 g monofilament testing have similarly acceptable levels of reliability and the neurothesiometer is the most reliable method of assessing vibration perception function. Use of a graduated tuning fork was slightly more reliable than other methods of tuning fork application however all had substantial reliability. Years of clinical experience only marginally affected test reliability overall and due to subjective nature of the tests we suggest that testing should be performed regularly and repetitively.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6966840PMC
http://dx.doi.org/10.1186/s13047-020-0371-9DOI Listing

Publication Analysis

Top Keywords

vibration perception
16
tuning fork
16
protective sensation
12
perception protective
8
people diabetes
8
inter- intra-rater
8
intra-rater reliability
8
clinical experience
8
graduated tuning
8
male 100%
8

Similar Publications

How to determine hands' vibration perception thresholds - a systematic review.

Behav Res Methods

December 2024

Algoritmi Research Centre, University of Minho, Campus de Azurém, 4800-058, Guimarães, Portugal.

The vibration perception threshold (VPT) is the minimum amplitude required for conscious vibration perception. VPT assessments are essential in medical diagnostics, safety, and human-machine interaction technologies. However, factors like age, health conditions, and external variables affect VPTs.

View Article and Find Full Text PDF

This study develops biomimetic strategies for slip prevention in prosthetic hand grasps. The biomimetic system is driven by a novel slip sensor, followed by slip perception and preventive control. Here, we show that biologically inspired sensorimotor pathways can be restored between the prosthetic hand and users.

View Article and Find Full Text PDF

Mental imagery is a crucial cognitive process, yet its underlying neural mechanisms remain less understood compared to perception. Furthermore, within the realm of mental imagery, the somatosensory domain is particularly underexplored compared to other sensory modalities. This study aims to investigate the influence of tactile imagery (TI) on cortical somatosensory processing.

View Article and Find Full Text PDF

Background: There is limited information on the immediate effects of whole-body vibration (WBV) on the upper limb. This study aims to determine the immediate effects of WBV on reaction speed and proprioception in young adult students' upper extremities.

Methods: In total, 62 students participated in the study.

View Article and Find Full Text PDF

Haptic devices typically rely on rigid actuators and bulky power supply systems, limiting wearability. Soft materials improve comfort, but careful distribution of stiffness is required to ground actuation forces and enable load transfer to the skin. We present Haptiknit, an approach in which soft, wearable, knit textiles with embedded pneumatic actuators enable programmable haptic display.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!