Background: Historically, there has been excellent survivorship of total hip arthroplasty with newer implants expected to outperform prior generations. Review of our institutional database identified problems with one particular cross-linked polyethylene liner in mid-term follow-up, which has not been reported in the literature. We report on the technical aspects of this liner, as well as the clinical and radiographic factors associated with early failure of this bearing in our practice.

Methods: A retrospective review of our institutional database was performed from January 2009 to June 2019 of patients who presented with significant osteolysis in the setting of prior total hip arthroplasty with a contemporary polyethylene liner. Demographic and radiographic measures were collected for all patients.

Results: Twelve patients were identified, with nine undergoing revision surgery. All patients demonstrated radiographic osteolysis at the time of revision surgery. The average time to diagnosis of failure was 55.9 months (range: 12-120 months). Average cup position preoperatively was 44 degrees inclination (range: 36-53 degrees) and 21 degrees anteversion (range: 10-34 degrees). Nine patients underwent revision of the acetabular component, with two also undergoing femoral component revision. The Manufacturer and User Facility Device Experience database revealed 22 reported cases of wear-related failure from 2009 to 2019.

Conclusions: Considering that no identifiable risk factors related to patient demographics or implant position were identified, the Exactech Connexion GXL liner may be prone to a high rate of early failure from wear and severe secondary osteolysis. We recommend close surveillance of patients with this bearing surface.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.arth.2019.12.043DOI Listing

Publication Analysis

Top Keywords

total hip
12
hip arthroplasty
8
review institutional
8
institutional database
8
polyethylene liner
8
early failure
8
revision surgery
8
failure
5
patients
5
early polyethylene
4

Similar Publications

Background: The angiotensin-converting enzyme (ACE) insertion/deletion (I/D) polymorphism (rs4340) is associated with the pathogenesis of heart failure (HF). This polymorphism may contribute to a greater propensity for severe HF and excess weight.

Objective: To evaluate adiposity, cardiac function, and their association with ACE I/D polymorphism in HF patients.

View Article and Find Full Text PDF

Surgical site infection (SSI) after total hip and knee arthroplasty (THA/TKA) is a major complication leading to morbidity and mortality. Perioperative irrigation, frequently with antiseptic compounds including povidone-iodine (PI), is the standard of care in reducing SSI. Evidence supporting the value of PI versus nonantiseptic substances varies.

View Article and Find Full Text PDF

Background: Currently, there is no clear standard for the surgical options for Evans-Jensen Type IV intertrochanteric femur fractures in elderly patients over 65 years old. This study aims to retrospectively analyze and compare the early postoperative limb function and quality of life of patients treated with total hip arthroplasty (THA) and proximal femoral nail antirotation (PFNA) for this type of fracture. We hypothesize that there is no significant difference in complications and postoperative recovery between the two surgical methods.

View Article and Find Full Text PDF

Background: The debate continues on whether combining core decompression (CD) with regenerative therapy provides a more effective treatment for early femoral head necrosis than CD alone. This systematic review and meta-analysis endeavored to assess its efficacy.

Methods: We systematically searched PubMed, Web of Science, and Cochrane Library through July 2024 for RCTs and cohort studies evaluating the impact of core decompression (CD) with regenerative therapy versus CD alone in early-stage osteonecrosis (ARCO I, II or IIIa or Ficat I or II) of the femoral head (ONFH).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!