Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: This study aimed to create and validate an integrated data acquisition system for gauging the force distribution between a laryngoscope and soft-tissue during trans-oral surgery.
Methods: Sixteen piezoresistive force sensors were interfaced to a laryngoscope and custom maxillary tooth guard. A protocol for calibrating the laryngoscope and maxilla sensors was developed using a motor-controlled linear stage and force measurements were validated against a digital scale. The system was initially tested during suspension laryngoscopy on three cadaver heads mounted on a cadaver head-holder. Intraoperative data was also collected from three patients undergoing head and neck tumor resection.
Results: Mean calibration error of the scope sensors was less than 150 g (n = 3) and mean maxilla sensor error was less than 200 g (n = 3). Peak scope mag-forces of 8.09 ± 6.61 kg and peak maxilla forces of 7.62 ± 4.57 kg were experienced during the cadaver trials. The peak scope sensor mag-force recorded during the intraoperative cases was 24.7 ± 4.53 kg, and the peak maxilla force was 22.0 ± 4.60 kg.
Conclusion: The data acquisition system was successfully able to record intraoperative force distribution data. The usefulness of this technology in informing surgeons during trans-oral surgery should be further evaluated in patients with varying anatomic and procedural characteristics.
Significance: Creation of a low-cost, integrated force-sensing system allows for the characterization of retraction forces at anatomic sites including the pharynx and larynx, brain, and abdomen. Real-time force detection provides surgeons with valuable intraoperative feedback and can be used to improve deformation models at various anatomic sites.