Objectives: Radio frequency (RF) pulses in magnetic resonance imaging (MRI) can interact with implanted devices and cause tissue damage. However, there are new devices that can safely perform measurements with liberal MRI conditions such as an RF transmission field B ≤ 2.0 μT. We investigated whether MRI in this case is limited for these technical reasons.

Methods: We selected typical MRI protocols of six body regions (brain, cervical spine, lumbar spine, knee, liver, heart) using two typical 1.5T MRI scanners. Overall, we adapted 62 sequences to B conditions and evaluated their diagnostic quality. For this, we measured signal-to-noise-ratio (SNR), contrast-to-noise-ratio (CNR), and geometric deviation (GD) as quality parameters, using phantom studies. For questionnaire studies, we selected pairs of original and adapted sequences in healthy volunteers. Blinded radiologists rated the images as single sequence rating and in direct comparison.

Results: Roughly one-third of the checked sequences were below the B limit. Here, 56 of the 62 adapted sequences showed at least the same image quality in single ratings. A reduction in SNR and/or CNR was found with 31 sequences and only one sequence with considerably increased GD. Especially, sequences with original high B values, PD sequences, and sequences of the Siemens knee and heart protocol were difficult to adapt, whereas most TSE and IR sequences had no clinical limitations.

Conclusion: By limiting the transmission field to B ≤ 2.0 μT, clinically relevant MR sequences can be adapted with nearly no reduction in image quality. Despite limiting the transmission field, high-quality MR imaging is possible. We could derive strategies for adaptation.

Key Points: • Despite limiting the transmission field, high-quality MRI is possible. • We could derive strategies for adapting the sequences to B≤ 2.0 μT. • This enables high-quality MRI of different body regions for patients with AD.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-019-06599-6DOI Listing

Publication Analysis

Top Keywords

transmission field
16
image quality
12
adapted sequences
12
limiting transmission
12
sequences
11
mri
8
field b ≤ 20 μt
8
body regions
8
despite limiting
8
field high-quality
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!