Guidance means accuracy: A randomized clinical trial on freehand versus guided dental implantation.

Clin Oral Implants Res

Department of Oral and Maxillofacial Surgery, Faculty of Medicine, University of Szeged, Szeged, Hungary.

Published: May 2020

Objectives: A randomized clinical trial was conducted to compare all three known static guided surgery protocols (pilot, partial, and full) with each other and with freehand surgery in terms of accuracy, under the same conditions.

Material And Methods: A total of 207 implants of the same brand and type were placed in 101 partially edentulous volunteers in need of implantation in the mandible or maxilla or both. All cases were digitally planned, and the comparison of the planned and actual implant positions was performed using a medical image analysis software with dedicated algorithms. The primary outcome variable was angular deviation (AD, degrees). The secondary outcome variables were coronal global deviation (CGD, mm), apical global deviation (AGD, mm), and voxel overlap (VO, %).

Results: AD showed stepwise improvement in significant steps as the amount of guidance increased. The highest mean AD (7.03° ± 3.44) was obtained by freehand surgery and the lowest by fully guided surgery (3.04° ± 1.51). As for the secondary outcome variables, all guided protocols turned out to be significantly superior to freehand surgery, but they were not always significantly different from each other.

Conclusions: As for the comparison that this study sought to perform, it can be said that the static guided approach significantly improves the accuracy of dental implant surgery as compared to freehand surgery. Furthermore, the results suggest that any degree of guidance yields better results than freehand surgery and that increasing the level of guidance increases accuracy.

Download full-text PDF

Source
http://dx.doi.org/10.1111/clr.13578DOI Listing

Publication Analysis

Top Keywords

freehand surgery
20
randomized clinical
8
clinical trial
8
static guided
8
surgery
8
guided surgery
8
secondary outcome
8
outcome variables
8
global deviation
8
freehand
6

Similar Publications

Ventriculoperitoneal shunt (VPS) insertion is a neurosurgical procedure done routinely for managing hydrocephalus. However, the technique of shunt insertion remains controversial. In this study, we retrospectively compared the accuracy of shunt placement using ultrasound (US) guidance to freehand insertion.

View Article and Find Full Text PDF

Objective: Mixed-reality (MR) applications provide opportunities for technical rehearsal, education, and estimation of surgical performance without the risk of patient harm. In this study, the authors provide a structured literature review on the current state of MR applications and their effects on neurosurgery training. They also introduce an MR prototype for neurosurgical spine training.

View Article and Find Full Text PDF

Accuracy of robot-assisted implant surgery versus freehand placement: a retrospective clinical study.

Int J Implant Dent

January 2025

Key Laboratory of Shaanxi Province for Craniofacial Precision Medicine Research, College of Stomatology, Xi'an Jiaotong University, Xi'an, China.

Purpose: This study evaluated the accuracy of implant placement using a robotic system (Remebot) compared to freehand surgery and explored factors influencing accuracy.

Methods: This retrospective study included 95 implants placed in 65 patients, divided into robot-assisted (50 implants) and freehand (45 implants) groups. Platform, apical, and angular deviations were measured by superimposing preoperative plans and the postoperative CBCT images.

View Article and Find Full Text PDF

Background: Virtual surgical planning (VSP) is an emerging method in head and neck reconstruction with demonstrated benefits, however, its economic viability is supported with mixed evidence.

Methods: A structured search was performed in five electronic databases. Studies that performed an economic evaluation on VSP in head and neck reconstruction were included.

View Article and Find Full Text PDF

Background: We investigated the perioperative outcome and oncologic safety of performing nipple-sparing mastectomy (NSM) through a single axillary incision (NSM-SAI) compared with performing NSM through a conventional incision (NSM-C).

Methods: We retrospectively reviewed 725 patients who underwent NSM for breast cancer between January 2010 and December 2023; 333 patients who underwent NSM with immediate reconstruction (IR) were enrolled. Surgical outcomes and oncologic outcomes of NSM-C (n = 184), NSM performed through SAI with a freehand approach (NSM-SAI-F; n = 92), and with an endoscopic approach (NSM-SAI-E; n = 57) were demonstrated.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!