Purposes: Multileaf collimator (MLC) positional accuracy during dynamic intensity modulation radiotherapy (IMRT) delivery is crucial for safe and accurate patient treatment. The deviations of individual leaf positions from its intended positions can lead to errors in the dose delivered to the patient and hence may adversely affect the treatment outcome. In this study, we propose a state-of-the-art machine learning (ML) method based on an artificial neural network (ANN) for accurately predicting the MLC leaf positional deviations during the dynamic IMRT treatment delivery priori using log file data.

Methods: Data of ten patients treated with sliding window dynamic IMRT delivery were retrospectively retrieved from a single-institution database. The patients' plans were redelivered with no patient on the couch using a Varian linear accelerator equipped with a Millennium 120 HD MLC system. Then the machine recorded log files data, a total of over 400 files containing 360 800 control points, were collected. A total of 14 parameters were extracted from the planning data in the log files such as leaf planned positions, dose fraction, leaf velocity, leaf moving status, leaf gap, and others. Next, we developed a feed-forward ANN architecture mapping the input parameters with the output to predict the MLC leaf positional deviations during the delivery priori. The proposed model was trained on 70% of the total data using the delivered leaf positional data as a target response. The trained model was then validated and tested on 30% of the available data. The model accuracy was evaluated using the mean squared error (MSE), regression plot, and error histogram.

Results: The deviations between the individual MLC planned and delivered positions can reach up to a few millimeters, with a maximum deviation of 1.2 mm. The predicted leaf positions at control points closely matched the delivered positions for all MLC leaves during the treatment delivery. The ANN model achieved a maximum MSE of 0.0001 mm (root MSE of 0.0097 mm) in predicting the leaf positions at control points of test data for each leaf. The correlation coefficient, that measures the goodness of fit, was perfect (R = 0.999) in all plots indicating an excellent agreement between the predicted and delivered MLC positions for the training, validation, and test data.

Conclusions: We successfully demonstrated a proposed ANN-based method capable of accurately predicting the individual MLC leaf positional deviations during the dynamic IMRT delivery priori. Our ML model based on ANN outperformed the reported accuracy in the literature of various ML models. The results of this study could be extended to actual application in the dose calculation/optimization, hence enhancing the gamma passing rate for patient-specific IMRT quality assurance.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.14014DOI Listing

Publication Analysis

Top Keywords

positional deviations
16
dynamic imrt
16
imrt delivery
16
delivery priori
16
leaf positional
16
deviations dynamic
12
leaf
12
leaf positions
12
mlc leaf
12
control points
12

Similar Publications

The title compound, CHNO, was synthesized by S2 reaction of bromo-methyl coumarin with 4,4-di-methyl-piperidine-2,6-dione. The mol-ecule crystalizes in the monoclinic system with space group 2/. The coumarin unit is almost planar with a dihedral angle between the aromatic rings of 0.

View Article and Find Full Text PDF

Robotic-assisted dental implantation represents a transformative innovation in modern dentistry, offering enhanced surgical precision and reduced variability. Despite its clinical adoption, the impact of anatomical and bone-related factors on placement accuracy remains underexplored. This retrospective study evaluated 54 implants placed in 30 patients using cone-beam computed tomography (CBCT) and virtual planning software to analyze deviations in crown position, apex position, and angulation.

View Article and Find Full Text PDF

In the last decades, maxillomandibular reconstruction has been revolutionised by the use of free flaps and virtual surgical planning technologies. However, the currently available applied physical cutting guides provide no intraoperative flexibility, and adjustments based on intraoperative findings are not possible. A novel augmented reality (AR)-guided technique is presented that allows for quick intraoperative surgical planning adaptations.

View Article and Find Full Text PDF

In vitro evaluation of six intraoral scanners for cone beam computed tomography data registration in robotic-assisted implant surgery: impact of scanning experience.

J Dent

March 2025

Experimental Center for Stomatological Engineering, Jiamusi University &Key Laboratory of Oral Biomedical Materials and Clinical Application, Jiamusi, Heilongjiang Province, P.R. China; Department of Digital Dental Implant, Affliated Stomatological Hospital of Jiamusi University & Stomatology College of Jiamusi University, Jiamusi, Heilongjiang Province, P.R. China. Electronic address:

Objectives: This study evaluated the accuracy of intraoral scanning (IOS) and cone beam computed tomography (CBCT) data registration (ICR) in robotic computer-assisted implant surgery (r-CAIS) using six different intraoral scanners, considering the impact of surgeon experience.

Methods: A total of 112 standardized mandibular partially edentulous models were assigned to six experimental groups and one control group based on the intraoral scanner used. In the control group, preoperative CBCT data were registered with CBCT data from a model containing an optical tracking marker (OT-Marker).

View Article and Find Full Text PDF

Background And Aims: Following an ankle sprain, clinical examination often reveals ankle dorsiflexion pain, which has been implicated as a risk factor for recurrent ankle sprains; however, the mechanism of ankle dorsiflexion pain has not been explored. Using magnetic resonance imaging (MRI), we evaluated the relationship between the limited dorsiflexion range of motion due to pain and the position of the talus. We investigated whether an anterior talofibular ligament (ATFL) tear affected ankle dorsiflexion.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!