Purpose: In the previous treatment planning system (TPS) for CyberKnife (CK), multileaf collimator (MLC)-based treatment plans could be created only by using the finite-size pencil beam (FSPB) algorithm. Recently, a new TPS, including the FSPB with lateral scaling option (FSPB+) and Monte Carlo (MC) algorithms, was developed. In this study, we performed basic and clinical end-to-end evaluations for MLC-based CK tumor-tracking radiotherapy using the MC, FSPB+, and FSPB.
Methods: Water- and lung-equivalent slab phantoms were combined to obtain the percentage depth dose (PDD) and off-center ratio (OCR). The CK M6 system and Precision TPS were employed, and PDDs and OCRs calculated by the MC, FSPB+, and FSPB were compared with the measured doses obtained for 30.8 × 30.8 mm and 60.0 × 61.6 mm fields. A lung motion phantom was used for clinical evaluation and MLC-based treatment plans were created using the MC. The doses were subsequently recalculated using the FSPB+ and FSPB, while maintaining the irradiation parameters. The calculated doses were compared with the doses measured using a microchamber (for target doses) or a radiochromic film (for dose profiles). The dose volume histogram (DVH) indices were compared for all plans.
Results: In homogeneous and inhomogeneous phantom geometries, the PDDs calculated by the MC and FSPB+ agreed with the measurements within ±2.0% for the region between the surface and a depth of 250 mm, whereas the doses calculated by the FSPB in the lung-equivalent phantom region were noticeably higher than the measurements, and the maximum dose differences were 6.1% and 4.4% for the 30.8 × 30.8 mm and 60.0 × 61.6 mm fields, respectively. The maximum distance to agreement values of the MC, FSPB+, and FSPB at the penumbra regions of OCRs were 1.0, 0.6, and 1.1 mm, respectively, but the best agreement was obtained between the MC-calculated curve and measurements at the boundary of the water- and lung-equivalent slabs, compared with those of the FSPB+ and FSPB. For clinical evaluations using the lung motion phantom, under the static motion condition, the dose errors measured by the microchamber were -1.0%, -1.9%, and 8.8% for MC, FSPB+, and FSPB, respectively; their gamma pass rates for the 3%/2 mm criterion comparing to film measurement were 98.4%, 87.6%, and 31.4% respectively. Under respiratory motion conditions, there was no noticeable decline in the gamma pass rates. In the DVH indices, for most of the gross tumor volume and planning target volume, significant differences were observed between the MC and FSPB, and between the FSPB+ and FSPB. Furthermore, significant differences were observed for lung D , V , and V between the MC, FSPB+, and FSPB.
Conclusions: The results indicate that the doses calculated using the MC and FSPB+ differed remarkably in inhomogeneous regions, compared with the FSPB. Because the MC was the most consistent with the measurements, it is recommended for final dose calculations in inhomogeneous regions such as the lung. Furthermore, the sufficient accuracy of dose delivery using MLC-based tumor-tracking radiotherapy by CK was demonstrated for clinical implementation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/mp.14013 | DOI Listing |
Front Oncol
October 2023
Department of Radiation Oncology and Physics, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, China.
Purpose: The accuracy of dose calculation is the prerequisite for CyberKnife (CK) to implement precise stereotactic body radiotherapy (SBRT). In this study, CK-MLC treatment planning for early-stage non-small cell lung cancer (NSCLC) were compared using finite-size pencil beam (FSPB) algorithm, FSPB with lateral scaling option (FSPB_LS) and Monte Carlo (MC) algorithms, respectively. We concentrated on the enhancement of accuracy with the FSPB_LS algorithm over the conventional FSPB algorithm and the dose consistency with the MC algorithm.
View Article and Find Full Text PDFJ Cancer
September 2023
Department of Radiation Oncology, Tianjin Medical University Cancer Institute and Hospital, National Clinical Research Center for Cancer, Tianjin's Clinical Research Center for Cancer, Key Laboratory of Cancer Prevention and Therapy, Tianjin, 300060, People's Republic of China.
The dose distribution in different optimization algorithm plans of stereotactic radiotherapy (SBRT) for locally advanced pancreatic cancer (LAPC) were compared and analyzed using monte carlo dose calculate algorithm (MC). A retrospective study analyzed 26 LAPC patients treated with SBRT. The SBRT plans were designed by raytracing (RT) and fine size pencil beam (FSPB) algorithms in the CyberKnife (CK) precision system, all of which met the requirements of clinical target dose and organ at risk (OAR).
View Article and Find Full Text PDFJ Med Phys
March 2020
Department of Physics, Florida Atlantic University, Boca Raton, Florida, USA.
Purpose: InCise™ multileaf collimator (MLC) was introduced for CyberKnife (CK) Robotic Radiosurgery System (CK-MLC) in 2015, and finite size pencil beam (FSPB) was the only available dose computation algorithm for treatment plans of CK-MLC system. The more advanced Monte Carlo (MC) dose calculation algorithm of lnCise™ was initially released in 2017 for the CK Precision™ treatment planning system (TPS) (v1.1) with new graphic processing unit (GPU) platform.
View Article and Find Full Text PDFMed Phys
March 2020
Department of Radiology, Nagoya University Graduate School of Medicine, Nagoya, Aichi, 466-8550, Japan.
Purpose: In the previous treatment planning system (TPS) for CyberKnife (CK), multileaf collimator (MLC)-based treatment plans could be created only by using the finite-size pencil beam (FSPB) algorithm. Recently, a new TPS, including the FSPB with lateral scaling option (FSPB+) and Monte Carlo (MC) algorithms, was developed. In this study, we performed basic and clinical end-to-end evaluations for MLC-based CK tumor-tracking radiotherapy using the MC, FSPB+, and FSPB.
View Article and Find Full Text PDFPhys Med
November 2018
European Cyberknife Center Munich, Munich, Germany; University Hospital of Cologne, Department for Stereotaxy and Functional Neurosurgery, Cologne, Germany.
Purpose: At introduction in 2014, dose calculation for the first MLC on a robotic SRS/SBRT platform was limited to a correction-based Finite-Size Pencil Beam (FSPB) algorithm. We report on the dosimetric accuracy of a novel Monte Carlo (MC) dose calculation algorithm for this MLC, included in the Precision™ treatment planning system.
Methods: A phantom was built of one slab (5.