study of glenoid perforation during total shoulder arthroplasty: the effects on stress & micromotion.

Comput Methods Biomech Biomed Engin

Department of Biomedical Engineering, School of Biomedical Engineering, Faculty of Engineering, Universiti Teknologi Malaysia, Skudai, Johor, Malaysia.

Published: April 2020

Glenoid perforation is not the intended consequence of the surgery and must be avoided. The analysis on biomechanical aspect of glenoid vault perforation remains unknown. The purpose of this study is to determine the impact of glenoid perforation towards stress distribution and micromotion at the interfaces. Eight glenoid implant models had been constructed with various size, number and type of fixation. A load of 750 N was applied to centre, superior-anterior and superior-posterior area. Implant perforation had minimal impact on stress distribution and micromotion at the interfaces. However, cement survival rate for implant without perforation was the highest with a difference of up to 37% compared to other perforated models. Besides that, implant fixation and high stresses at the implant had more of an impact on implant instability than implant perforation. As a conclusion, glenoid perforation did not influence the stress distribution and micromotion, but, it reduced cement survival rate and increase the stress critical volume.

Download full-text PDF

Source
http://dx.doi.org/10.1080/10255842.2019.1709828DOI Listing

Publication Analysis

Top Keywords

glenoid perforation
16
stress distribution
12
distribution micromotion
12
implant perforation
12
perforation
8
micromotion interfaces
8
cement survival
8
survival rate
8
implant
7
stress
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!