Objective: A phantom for diffusion-weighted imaging is required to standardize quantitative evaluation. The objectives were to develop a phantom simulating various cell densities and to evaluate repeatability.

Materials And Methods: The acrylic fine particles with three different diameters were used to simulate human cells. Four-degree cell density components were developed by adjusting the volume of 10-μm particles (5, 20, 35, and 50% volume, respectively). Two-degree components to simulate cell edema were also developed by adjusting the diameter without changing number (17% and 40% volume, respectively). Spearman's rank correlation coefficient was used to find a significant correlation between apparent diffusion coefficient (ADC) and particle density. Coefficient of variation (CV) for ADC was calculated for each component for 6 months. A p value < 0.05 represented a statistically significance.

Results: Each component (particle ratio of 5, 17, 20, 35, 40, and 50% volume, respectively) presented ADC values of 1.42, 1.30, 1.30, 1.12, 1.09, and 0.89 (× 10 mm/s), respectively. A negative correlation (r =  - 0.986, p < 0.05) was observed between ADC values and particle ratio. CV for ADC was less than 5%.

Discussion: A phantom simulating the diffusion restriction correlating with cell density and size could be developed.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s10334-019-00823-6DOI Listing

Publication Analysis

Top Keywords

phantom simulating
8
cell edema
8
developed adjusting
8
cell
5
development phantom
4
simulating extracellular
4
extracellular space
4
space tumor
4
tumor cell
4
cell growth
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!