Background: Numerous tests have been proposed to evaluate sperm DNA integrity. To assess the sperm chromatin dispersion (SCD) test in an andrology laboratory, twenty-five men attending Clermont-Ferrand (France) University Hospital's Center for Reproductive Medicine were recruited. Sperm DNA damage was measured in the same semen samples using the SCD test and the Terminal Uridine Nick-end Labeling by flow cytometry technique (TUNEL/FCM) after density gradient centrifugation.

Results: SCD test reliability between readings, readers or slides was clearly established with very high agreement between measurements (Intraclass correlation coefficient (ICC) at 0.97, 0.95 and 0.98 respectively). Despite very good agreement between the SCD test and TUNEL/FCM (ICC at 0.94), the SCD test tended to slightly but significantly underestimate DNA damage compared with TUNEL ( = 0.0127). This systematic difference between the two techniques was - 3.39 ± 1.45% (mean ± SE).

Conclusions: Andrology laboratories using the SCD test to measure sperm DNA damage need to know that it appears to give slightly underestimated measurements compared to TUNEL/FCM. However, this systematic underestimation is very small in amplitude. Both techniques give almost perfectly congruent results. Our study underlines the importance for each laboratory to validate its method to assess sperm DNA damage before implementing it in routine andrology lab practice.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6933933PMC
http://dx.doi.org/10.1186/s12610-019-0098-2DOI Listing

Publication Analysis

Top Keywords

scd test
28
sperm dna
20
dna damage
20
test terminal
8
terminal uridine
8
uridine nick-end
8
nick-end labeling
8
labeling flow
8
flow cytometry
8
cytometry technique
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!