Purpose: Corneal opacity is a leading cause of reversible blindness worldwide. An electronic corneal prosthesis, or intraocular projector, could potentially restore high-quality vision without need for corneal clarity.

Materials And Methods: Four intraocular projection systems were constructed from commercially available electronic components and encased in biocompatible plastic housing. They were tested for optical properties, biocompatibility, heat dissipation, waterproofing, and accelerated wear. A surgical implantation technique was developed.

Results: Intraocular projectors were produced of a size that can fit within the eye. Their optics produce better than 20/200 equivalent visual acuity. MTT assay demonstrated no cytotoxicity of devices . Temperature testing demonstrated less than 2°C increase in temperature after 1 h. Three devices lasted over 12 weeks under accelerated wear conditions. Implantation surgery was demonstrated via corneal trephination insertion in a cadaver eye.

Conclusion: This is the first study to demonstrate and characterize fully functional intraocular projection systems. This technology has the potential to be an important new tool in the treatment of intractable corneal blindness.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9997130PMC
http://dx.doi.org/10.1080/02713683.2019.1708957DOI Listing

Publication Analysis

Top Keywords

electronic corneal
8
corneal prosthesis
8
intraocular projection
8
projection systems
8
accelerated wear
8
corneal
6
characterization electronic
4
prosthesis system
4
system purpose
4
purpose corneal
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!