Comparison of oscillometry devices using active mechanical test loads.

ERJ Open Res

Dept of Anesthesiology and Intensive Therapy, Semmelweis University, Budapest, Hungary.

Published: October 2019

Noninvasiveness, low cooperation demand and the potential for detailed physiological characterisation have promoted the use of oscillometry in the assessment of lung function. However, concerns have been raised about the comparability of measurement outcomes delivered by the different oscillometry devices. The present study compares the performances of oscillometers in the measurement of mechanical test loads with and without simulated breathing. Six devices (five were commercially available and one was custom made) were tested with mechanical test loads combining resistors (), gas compliances () and a tube inertance (), to mimic respiratory resistance ( ) and reactance ( ) spectra encountered in clinical practice. A ventilator was used to simulate breathing at tidal volumes of 300 and 700 mL at frequencies of 30 and 15 min, respectively. Measurements were evaluated in terms of , , , resonance frequency ( ), reactance area ( ) and resistance change between 5 and 20 or 19 Hz ( ). Increasing test loads caused progressive deviations in and from calculated values at various degrees in the different oscillometers. While mean values of were recovered acceptably, some devices exhibited serious distortions in the frequency dependences of and , leading to large errors in , , , and . The results were largely independent of the simulated breathing. Simplistic calibration procedures and mouthpiece corrections, in addition to unknown instrumental and signal processing factors, may be responsible for the large differences in oscillometry measures. Rigorous testing and ongoing harmonisation efforts are necessary to better exploit the diagnostic and scientific potential of oscillometry.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6926364PMC
http://dx.doi.org/10.1183/23120541.00160-2019DOI Listing

Publication Analysis

Top Keywords

test loads
16
mechanical test
12
oscillometry devices
8
simulated breathing
8
comparison oscillometry
4
devices
4
devices active
4
active mechanical
4
test
4
loads
4

Similar Publications

Assessment Implementation Measures (AIM): The Journey Towards the Tool Validity.

J Coll Physicians Surg Pak

January 2025

Department of Oral and Maxillofacial Surgery, Frontier Medical and Dental College, Abbottabad, Pakistan.

Objective: To establish the construct validity of the Assessment Implementation Measures (AIM) tool to accurately assess faculty perspectives on implemented assessment systems, facilitating the alignment with set standards.

Study Design: Qualitative Study. Place and Duration of the Study: The study was carried out at RIPHAH International University and data were collected from participants (senior faculty members) involved in teaching and assessment of undergraduates from various medical and dental colleges in Pakistan.

View Article and Find Full Text PDF

Objective: To test the short and long-term effects of consuming carbohydrate-rich beverages on patient-centred outcomes after caesarean delivery under spinal anaesthesia.

Study Design: A prospective randomised controlled study. Place and Duration of the Study: Department of Obstetrics and Gynaecology, Karaman Training and Research Hospital, Karaman, Turkiye, between May 2023 and February 2024.

View Article and Find Full Text PDF

Glassphalt suffers from performance defects, especially against moisture damage and fatigue cracking. In this research, the performance of glassphalt modified with CF has been evaluated against moisture damage, fatigue cracking and rutting. Based on this, Modified Lottman, Wilhelmy Plate (WP), Indirect Tensile Stiffness Modulus (ITSM), Indirect Tensile Fatigue (ITF), and Repeated Load Axial (RLA) tests have been performed on glassphalt modified with CF.

View Article and Find Full Text PDF

Introduction: With the increased use of CTs in cases with trimalleolar ankle fractures, bone fragments between the posterior malleolus and the rest of the articular surface tibial plafond surface - described as intercalary fragments (ICFs) - can be recognized. The aim of this study was to determine the ICF size threshold for a significant change in the pressure distribution at the ankle joint, having a considerable impact on the remaining cartilage of the joint.

Design And Methods: Eight human cadaveric lower legs were used, and a posterior malleolus Bartonicek II fracture was created with sequential 2mm, 4mm, 6mm and 8mm ICFs.

View Article and Find Full Text PDF

A data-driven framework for developing a unified density-modulus relationship for the human lumbar vertebral body.

J Mech Behav Biomed Mater

January 2025

Department of Mechanical Engineering, Boston University, Boston, MA 02215, USA; Center for Multiscale and Translational Mechanobiology, Boston University, Boston, MA 02215, USA; Department of Biomedical Engineering, Boston University, Boston, MA 02215, USA.

Despite the broad agreement that bone stiffness is heavily dependent on the underlying bone density, there is no consensus on a unified relationship that applies to both cancellous and cortical compartments. Bone from the two compartments is generally assessed separately, and few mechanical test data are available for samples from the transitional regions between them. In this study, we present a data-driven framework integrating experimental testing and numerical modeling of the human lumbar vertebra through an energy balance criterion, to develop a unified density-modulus relationship across the entire vertebral body, without the necessity of differentiation between trabecular and cortical regions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!