Background: Hollow, bendable manipulators can advance implant delivery in minimally invasive surgery, by circumventing the drawbacks of straight-line delivery and fostering single-port approaches. Variable stiffness manipulators are sought to be safe and effective.

Methods: We designed and experimentally assessed a cable-driven articulated/continuum manipulator, devised for cardiac valve delivery. Positioning and stiffening were teleoperated, based on cable shortening. Stiffening was parameterized by using the leading tension (LT, ie, tension of the cables driving bending). We assessed positioning (repeatability/reversibility along eight/two bending directions) and stiffening (eight bent configurations).

Results: We achieved good repeatability and reversibility (mean errors <1% and 1.5%, respectively, of the workspace characteristic length). Stiffening was effective (up to 9-fold increase, depending on pose). Stiffening was linearly correlated (R = 0.92) with LT for all the considered configurations.

Conclusion: We accurately positioned and effectively stiffened the manipulator in several bent configurations. The proposed stiffness modulation strategy can be extended to other manipulators.

Download full-text PDF

Source
http://dx.doi.org/10.1002/rcs.2072DOI Listing

Publication Analysis

Top Keywords

positioning stiffening
8
articulated/continuum manipulator
8
implant delivery
8
delivery minimally
8
minimally invasive
8
invasive surgery
8
stiffening articulated/continuum
4
manipulator implant
4
delivery
4
surgery background
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!