Introduction: Endometrial cancer (EC) is a leading cause of female cancer in developed countries. The total abdominal hysterectomy and bilateral salpingo-oophorectomy with pelvic lymph node dissection is the primary management of endometrial cancer. Afterwards vaginal brachytherapy can be preferred either alone or combined with external beam radiotherapy to avoid the vault recurrence. This study was to assess the in-house multichannel applicator for HDR vaginal brachytherapy and dosimetric compression with the single channel applicator through DVHs. The main objective of this study was to design and development of an in-house multichannel applicator for HDR vaginal brachytherapy and dosimetric compression with the single channel applicator through DVHs.

Materials And Methods: The multichannel applicator is a solid cylinder with 3 cm diameter, 13 cm length. It has a central channel surrounded by eight channels with a periphery arrangement in a single circle. We randomly selected eleven patients with endometrial carcinoma and 7Gy/per fraction of HDR dose was prescribed to CTV. Retrospectively, two 3D inverse treatment plans were created for each patient, with single channel loading and multichannel loading and the dose distribution of both plans could be compared. CTV coverage, rectum and bladder doses were compared.

Results: The DVH analysis showed statistically significant difference between single and multichannel plan, that is for D95 of CTV (p=0.008), D100 of CTV (p=0.004) and 2cc of CTV (p=0.003). The p value for 1cc, 2cc and V70 are 0.003, 0.003 and 0.003 for rectum. On the other hand, bladder DVHs showed large difference between single and multichannel plan yet it is not statistically significant, and the p values for 1cc, 2cc and V 80 are 0.012, 0.009 and 0.225.

Conclusion: The authors conclude that in house multichannel applicator with 3D inverse treatment planning techniques improves the dosimetric advantage over single channel applicators.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7173395PMC
http://dx.doi.org/10.31557/APJCP.2019.20.12.3805DOI Listing

Publication Analysis

Top Keywords

multichannel applicator
20
single channel
20
vaginal brachytherapy
16
in-house multichannel
12
applicator hdr
12
hdr vaginal
12
brachytherapy dosimetric
12
channel applicator
12
design development
8
development in-house
8

Similar Publications

Background And Purpose: In vivo dosimetry is not standard in brachytherapy and some errors go undetected. The aim of this study was to evaluate the accuracy of multi-channel vaginal cylinder pulsed dose-rate brachytherapy using in vivo dosimetry.

Materials And Methods: In vivo dosimetry data was collected during the years 2019-2022 for 22 patients (32 fractions) receiving multi-channel cylinder pulsed dose-rate brachytherapy.

View Article and Find Full Text PDF

Background: Endometrial cancer is one of the most common gynecological malignancies in the world. Vaginal brachytherapy is an important postoperative adjuvant treatment for endometrial cancer. However, a common problem with existing applicators is insufficient dose at the vaginal apex.

View Article and Find Full Text PDF

Objective: This study was conducted to compare the differences between 3D-printed multichannel non-co-planar vaginal applicators and single-channel vaginal applicators in cervical cancer patients with positive or close surgical margins.

Methods: Between January 2015 and June 2023, 104 cervical cancer patients who underwent radical surgery with positive or close surgical margins were enrolled to receive concurrent intensity-modulated chemoradiotherapy combined with 3D-printed multichannel non-co-planar vaginal applicators (3D-printed group, 41 patients) or single-channel vaginal applicators (single-channel group, 63 patients) guided brachytherapy. The dosimetric parameters, 5-year local control (LC), progression-free survival (PFS), overall survival (OS) of two groups were retrospectively analyzed.

View Article and Find Full Text PDF

Background And Purpose: Ir192 vaginal brachytherapy (IBT) is commonly used for patients with postoperative endometrial cancer (EC). We devised a novel multichannel vaginal applicator that could be equipped with an electronic brachytherapy (EBT) device. We aimed to explore the differences in physical parameters between the EBT and IBT.

View Article and Find Full Text PDF

Purpose: To compare the dosimetric performance of vaginal intensity-modulated brachytherapy (IM-BRT) applicator and single- (SC-BRT) and multi-channel brachytherapy (MC-BRT) applicators for vaginal cuff brachytherapy (VC-BRT).

Material And Methods: Fifteen patients with uterine-confined endometrium cancer who received adjuvant VC-BRT were included in this study. IM-BRT, SC-BRT, and MC-BRT treatment plans were created for two different clinical target volume (CTV) definitions: 1.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!