A von Hamos-type hard X-ray spectrometer at the PETRA III beamline P64.

J Synchrotron Radiat

Department Chemie and Center for Sustainable Systems Design (CSSD), Universität Paderborn, Warburger Straße 100, 33098 Paderborn, Germany.

Published: January 2020

The design and performance of the high-resolution wavelength-dispersive multi-crystal von Hamos-type spectrometer at PETRA III beamline P64 are described. Extended analyzer crystal collection available at the beamline allows coverage of a broad energy range from 5 keV to 20 keV with an energy resolution of 0.35-1 eV. Particular attention was paid to enabling two-color measurements by a combination of two types of analyzer crystals and two two-dimensional detectors. The performance of the spectrometer is demonstrated by elastic-line and emission-line measurements on various compounds.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6927517PMC
http://dx.doi.org/10.1107/S1600577519013638DOI Listing