Importance: Timely and complete disclosure of medical device defects is necessary to manage patient care safely and effectively.
Objectives: To determine if the manufacturer's recommendations following the recall of a medical device were timely and complete, the follow-up information and data provided to patients and physicians were adequate for managing patient care, and the actions taken by the US Food and Drug Administration (FDA) regarding the recall were appropriate.
Design, Setting, And Participants: This single-center retrospective case series included 90 of 448 patients who were implanted with a cardiac resynchronization therapy pacemaker at the Minneapolis Heart Institute from May 2003 through January 2011; this pacemaker was recalled in November 2015. In addition, returned product reports submitted by the manufacturer to the FDA via the Manufacturer and User Facility Device Experience (MAUDE) database from January 2008 through December 2018 were analyzed.
Main Outcomes And Measures: Clinical outcomes were serious adverse clinical events that occurred before and after the November 2015 recall notifying physicians and patients that the device's battery could fail unexpectedly because of high internal impedance. Technical outcomes were signs and causes of failure.
Results: Five of 90 patients observed during 2015 experienced syncope when their pacemakers stopped pacing owing to battery or wire connection defects prior to the recall. Of the 90 patients, 37 (41%) were men, and the median (interquartile range) age at implantation was 71.3 (66.1-78.2) years. Analysis of the MAUDE data revealed that battery failures prior to the recall were associated with serious adverse events that included 1 death, 1 cardiac arrest, 5 syncopal attacks, and 6 heart failure exacerbations; 3 additional prerecall syncopal events were caused by wire connection defects. The manufacturer and the FDA were aware of the battery and wire connection defects for 19 months before issuing the recall, yet the wire connection problem was not included in the advisory and physicians were not informed that interrogating the pacemaker could result in loss of pacing. The FDA classified the recall as class II rather than the more critical class I.
Conclusions And Relevance: This case series study of patients implanted with a defective pacemaker found that the pacemaker recall was delayed and that subsequent communications did not include all critical information needed for safe and effective patient care. These findings should prompt reforms in how the medical device industry and the FDA manage future medical device recalls.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6990820 | PMC |
http://dx.doi.org/10.1001/jamainternmed.2019.5171 | DOI Listing |
Hypertension
January 2025
Clinical Research Institute, Institute of Advanced Clinical Medicine, Peking University, Beijing, China (X.Z., W.X., Y.W.).
Background: Although the information on the validation status of electronic sphygmomanometer (ES) devices in use in health care institutions and households is much more clinically relevant than that of ES models available on the market, it remains insufficient.
Methods: A national survey was conducted across all administrative regions of mainland China to assess the validation status of ESs. Fifty-eight cities were selected with stratification by municipality, provincial capital, and other cities, and health care institutions and households in each city were chosen by convenience to identify ES devices in use according to the study protocol.
Heliyon
January 2025
BRITElab, Harry Perkins Institute of Medical Research, QEII Medical Centre Nedlands and Centre for Medical Research, The University of Western Australia, Perth, Australia.
Breast-conserving surgery accompanied by adjuvant radiotherapy is the standard of care for patients with early-stage breast cancer. However, re-excision is reported in 20-30 % of cases, largely because of close or involved tumor margins in the specimen. Several intraoperative tumor margin assessment techniques have been proposed to overcome this issue, however, none have been widely adopted.
View Article and Find Full Text PDFHeliyon
January 2025
Faculty of Medicine and Medical Sciences, University of Balamand, Kalhat, Lebanon.
Background: Transvenous pacemakers (TVP) and leadless pacemakers (LP) are two reliable permanent modalities for the treatment of heart rhythm disorders. Several observational studies explored the safety and efficacy of the two devices. The aim of this meta-analysis study is to present a comparative analysis of the safety of leadless versus transvenous pacemakers.
View Article and Find Full Text PDFJ Soc Cardiovasc Angiogr Interv
December 2024
Department of Cardiology, HonorHealth/Scottsdale Shea Medical Center, Scottsdale, Arizona.
Chronic total occlusion (CTO) percutaneous coronary intervention (PCI) is high risk compared to non-CTO PCI. Iatrogenic coronary artery hematoma formation is a common occurrence during CTO PCI, impairing true lumen visualization. We describe the use of a continuous mechanical suction (CMS) device in 2 applications in which it was used for successful subintimal hematoma decompression and distal vessel re-entry.
View Article and Find Full Text PDFRSC Adv
January 2025
Institute of Atomic and Molecular Sciences, Academia Sinica Taipei 106 Taiwan
Extreme ultraviolet (EUV) lithography is a cutting-edge technology in contemporary semiconductor chip manufacturing. Monitoring the EUV beam profiles is critical to ensuring consistent quality and precision in the manufacturing process. This study uncovers the practical use of fluorescent nanodiamonds (FNDs) coated on optical image sensors for profiling EUV and soft X-ray (SXR) radiation beams.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!