The Effect of Metal Instrumentation on Patient and Surgical Team Scatter Radiation Exposure Using Mini C-Arm in a Simulated Forearm Fracture Fixation Model.

J Am Acad Orthop Surg Glob Res Rev

Grandview Medical Center, Dayton, OH (Dr. Groover, Dr. Bamberger, Dr. Evans, and Dr. Gazaille); the Genesys Regional Medical Center, Grand Blanc, MI (Dr. Hinkley); and the Marian University College of Osteopathic Medicine, Indianapolis, IN (Dr. Gerow).

Published: June 2019

Introduction: To our knowledge, no studies have studied the effect of metal instrumentation (MI) in the field of fluoroscopy regarding changes in the intensity, direction of scatter, and degree of radiation exposure to the patient and surgical team. The goal of this study was to determine whether the presence of MI increases scatter radiation exposure to the patient and surgical team when using a mini C-arm in the horizontal and vertical positions.

Methods: Four trials were conducted using a lamb limb specimen and a mini C-arm to simulate a forearm/wrist fracture fixation scenario. Radiation scatter percentages were measured with the mini C-arm in a vertical and horizontal position with and without the presence of MI (a six-hole 3.5-mm limited-contact dynamic compression plate attached to the specimen with six cortical screws and a self-retaining retractor) using a parallel plate radiation detector.

Results: The patient, scrub technician, circulating nurse, and anesthesiologist were exposed to no detectable radiation. In the horizontal position with the presence of MI, there was a 181-fold increase in scatter radiation exposure to the first assistant's eyes (0.016% versus 2.893%, 1.4 × 10 Sv/min versus 3.5 × 10 Sv/min) and increased exposure to the surgeon's hands compared with the horizontal position with no MI. In the vertical position, the scatter radiation received by the first assistant's eyes increased (zero versus 2.9 × 10 Sv/min) with MI present, whereas the only radiation measured for the surgeon was in the right hand which did not change with MI present (2.2 × 10 Sv/min).

Discussion: MI in the field of fluoroscopy increases scatter radiation exposure to a degree that may place the first assistant's yearly eye exposure in excess of the International Commission on Radiological Protection limit. Surgeons and their assistants should wear lead aprons, thyroid shields, and leaded glasses and minimize the usage of fluoroscopy with MI in the field.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6917351PMC
http://dx.doi.org/10.5435/JAAOSGlobal-D-18-00089DOI Listing

Publication Analysis

Top Keywords

scatter radiation
20
radiation exposure
20
mini c-arm
16
patient surgical
12
surgical team
12
horizontal position
12
radiation
10
metal instrumentation
8
fracture fixation
8
field fluoroscopy
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!