Comparison of the Hounsfield unit in CT scan with the gray level in cone-beam CT.

J Dent Res Dent Clin Dent Prospects

Department of Oral and Maxillofacial Radiology, Faculty of Dentistry, Tabriz University of Medical Sciences, Tabriz, Iran.

Published: October 2019

The present study was undertaken to compare the Hounsfield Unit (HU) in computed tomography (CT) with the gray level in CBCT in human tissues. In this study, 25 different soft and hard tissues were evaluated in 21 patients. CBCT images were taken with Newtom VGi machine (Verona, Italy) and CT images were prepared with Somatom Sensation unit (Siemens, Germany). The HU values of soft and hard tissues were compared with the gray level values of CBCT images. There was a strong correlation between the HU in CT and the gray level in CBCT in soft tissues (P<0.001, R = 0.85) and hard tissues (P<0.001, R = 0.74) and in general (P<0.001, R = 0.91). A high degree of agreement was seen between HU in CT and gray level in CBCT in both hard and soft tissues. Since the gray level in CBCT was similar to HU in CT and can be used as a parameter determine bone density in implant treatment and also to determine the bone type, the CBCT technique is recommended in such cases due to its low radiation dose, short time and low cost compared to CT.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6904917PMC
http://dx.doi.org/10.15171/joddd.2019.028DOI Listing

Publication Analysis

Top Keywords

gray level
16
hounsfield unit
8
level cbct
8
soft hard
8
hard tissues
8
cbct images
8
comparison hounsfield
4
unit scan
4
gray
4
scan gray
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!