Surgical stapling devices are known for their reliability and convenience. A letter to health care professionals published by the US Food and Drug Administration in March 2019 highlighted the increasing number of adverse events associated with surgical staplers. Driven by a case of stapler malfunction during an elective laparoscopic sleeve gastrectomy, we performed a literature review to investigate the incidence of primary stapler malfunction. We also discuss the common types and an approach to its management. PubMed, MEDLINE, and EMBASE databases were searched for articles discussing surgical stapler malfunction. Twelve studies were selected that described the incidence and/or consequences of primary stapler malfunction. A narrative synthesis was performed. From observational studies, the incidence ranged from 0.022% to 2.3%. A prospective survey reported that 86% of laparoscopic surgeons either had personal experience with or knew of surgeons who experienced stapler malfunction, implying a higher incidence. Underreporting has been an issue as manufacturers can get exemptions from public reporting. Significantly, higher malfunctions have been reported after exemptions were lifted. The most common types of stapler malfunction are stapler misfire and stapler locking. Major morbidity and mortality have been reported. . Surgeons are increasingly reliant on technological innovations. Stapling failure occurs and it is imperative to be aware of this. Given the high volume of stapler use, a high percentage of surgeons are likely to encounter this problem in their career. It is important to have an approach to the prevention and management of this potentially catastrophic complication.

Download full-text PDF

Source
http://dx.doi.org/10.1177/1553350619889274DOI Listing

Publication Analysis

Top Keywords

stapler malfunction
28
primary stapler
12
stapler
10
literature review
8
incidence primary
8
common types
8
malfunction
7
incidence
5
review incidence
4
malfunction surgical
4

Similar Publications

Intraoperative robotic surgical system-related problems in robot-assisted thoracoscopic surgery.

Gen Thorac Cardiovasc Surg

September 2024

Department of Thoracic Surgery, Tokyo Women's Medical University, 8-1 Kawada-Cho, Shinjuku-Ku, Tokyo, 162-8666, Japan.

Background: Malfunctions of robotic instruments during robotic surgery are well known to occur; however, detailed reports on the inherent problems associated with robotic instruments and robotic surgical systems are scarce. The objective of this study was to retrospectively investigate the intraoperative problems associated with robotic surgical systems and robotic instruments.

Materials And Methods: This was a single-center retrospective study.

View Article and Find Full Text PDF

Background: Surgical staplers and clip appliers are commonly used and have a potential to malfunction, which may result in serious injury or death. These events are self-reported to the Food and Drug Administration and compiled in the Food and Drug Administration's Manufacturer and User Facility Device Experience database. This study characterizes mortality related to surgical stapler and clip applier failure reported in the Food and Drug Administration's Manufacturer and User Facility Device Experience database.

View Article and Find Full Text PDF

Background: Staple line leaks are a serious problem in bariatric surgery and a major cause of serious morbidity and mortality. Adverse events caused by medical devices are reported to the Food and Drug Administration which maintains the Manufacturer and User Facility Device Experience (MAUDE) database. We examined adverse stapler events reported to the MAUDE database, specifically with regards to bariatric surgery.

View Article and Find Full Text PDF

Robotic stapler use: Is it safe?-FDA database analysis across multiple surgical specialties.

PLoS One

November 2021

Department of Urology and Renal Transplantation, University of Toledo, Toledo, OH, United States of America.

Introduction: Robotic-assisted techniques are common across many surgical subspecialties. While robotic stapling offers increased surgeon control, there is limited information on surgical complications related to robotic stapler use.

Methods: We reviewed the FDA's MAUDE database for adverse events related to robotic stapler use.

View Article and Find Full Text PDF

Background: We compared surgical device malfunction reports in the Food and Drug Administration (FDA) public Manufacturer and User Facility Device Experience (MAUDE) with those in the FDA nonpublic Alternative Summary Reporting (ASR).

Methods: General surgery device product code categories in MAUDE and ASR from 1999 to 2018 were identified. Changes in the rates of categories and adverse events were evaluated by Poisson regression.

View Article and Find Full Text PDF