Purpose: With the increasing use of MR-guided radiation therapy (MRgRT), it becomes important to understand and explore accuracy of medical dosimeters in the presence of magnetic field. The purpose of this work is to characterize metal-oxide-semiconductor field-effect transistors (MOSFETs) in MRgRT systems at 0.345 T magnetic field strength.

Methods: A MOSFET dosimetry system, developed by Best Medical Canada for in-vivo patient dosimetry, was used to study various commissioning tests performed on a MRgRT system, MRIdian Linac. We characterized the MOSFET dosimeter with different cable lengths by determining its calibration factor, monitor unit linearity, angular dependence, field size dependence, percentage depth dose (PDD) variation, output factor change, and intensity modulated radiation therapy quality assurance (IMRT QA) verification for several plans. MOSFET results were analyzed and compared with commissioning data and Monte Carlo calculations.

Results: MOSFET measurements were not found to be affected by the presence of 0.345 T magnetic field. Calibration factors were similar for different cable length dosimeters either placed at the parallel or perpendicular direction to the magnetic field, with variations of less than 2%. The detector showed good linearity (R  = 0.999) for 100-600 MUs range. Output factor measurements were consistent with ionization chamber data within 2.2%. MOSFET PDD measurements were found to be within 1% for 1-15 cm depth range in comparison to ionization chamber. MOSFET normalized angular response matched thermoluminescent detector (TLD) response within 5.5%. The IMRT QA verification data for the MRgRT linac showed that the percentage difference between ionization chamber and MOSFET was 0.91%, 2.05%, and 2.63%, respectively for liver, spine, and mediastinum.

Conclusion: MOSFET dosimeters are not affected by the 0.345 T magnetic field in MRgRT system. They showed physics parameters and performance comparable to TLD and ionization chamber; thus, they constitute an alternative to TLD for real-time in-vivo dosimetry in MRgRT procedures.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6964768PMC
http://dx.doi.org/10.1002/acm2.12799DOI Listing

Publication Analysis

Top Keywords

magnetic field
20
ionization chamber
16
radiation therapy
12
0345 t magnetic
12
mosfet
9
mosfet dosimeter
8
mr-guided radiation
8
therapy mrgrt
8
mrgrt linac
8
mrgrt system
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!