Fibrin-d-dimer (d-dimer) is essential for the diagnosis and treatment of thrombosis patients. The new ACL TOP 50- series used for d-dimer determination, integrates a preanalytical interference check for hemolysis, icterus and lipemia (HIL-test). Using earlier versions of ACL TOP, HIL was evaluated by visual inspection by the biomedical laboratory scientist. With the new integrated HIL-test, lipemia is determined by measuring turbidity in the sample at 671 nm, reflecting the lipid content in the sample. Using the new ACL TOP 50- series, we observed more samples being rejected for d-dimer analysis due to lipemia-interference evaluated by the integrated HIL-test suggesting a discrepancy between the former triglyceride-based cut-off and the new turbidity-based cut-off. Therefore, to re-evaluate the lipemia interference cut-off with the turbidimetric measurement, we did a spike-in experiment, using intralipid as lipemia simulation. Three × six different patient pools were prepared from sodium-citrate plasma. We observed no interference in the d-dimer assay by intralipid concentrations resulting in milli absorbances (mAbs) below 3000 mAbs. Thus, the cut-off for lipemia interference using the integrated HIL-test on ACL TOP 550 should be 3000 mAbs. This cut-off will reduce the number of samples rejected due to false positive interference and at the same time reduce 'hands on' time for the biomedical laboratory scientists and minimize the risk of subjective evaluation by a visual inspection.

Download full-text PDF

Source
http://dx.doi.org/10.1080/00365513.2019.1703214DOI Listing

Publication Analysis

Top Keywords

acl top
20
lipemia interference
12
integrated hil-test
12
top 50-
8
50- series
8
visual inspection
8
biomedical laboratory
8
samples rejected
8
3000 mabs cut-off
8
interference
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!