Optimisation of the drill-in behaviour of the EcoFit® SC threaded cup.

Biomed Tech (Berl)

Department of Orthopaedics, Physical Medicine and Rehabilitation, University Hospital, LMU Munich, Marchioninistr. 15, 81377 Munich, Germany.

Published: August 2020

AI Article Synopsis

  • Threaded cups, specifically the EcoFit model, show good clinical outcomes but have had issues with incomplete placement in surgeries.
  • A study aimed to investigate the drill-in behaviour of the EcoFit cup compared to the SC cup, using foam blocks of varying densities and measuring factors such as primary stability and overhang after implanting.
  • Results indicated that the EcoFit cup had lower primary stability and could not be positioned as deeply as the SC cup under standard conditions; however, overreaming the implant cavity improved the stability and positioning of the EcoFit cup significantly.

Article Abstract

Threaded cups show good clinical results when implanted correctly. In clinical use, multiple cases with an incomplete placement of the EcoFit threaded cup (implantcast) were observed. This behaviour could not be explained intra- and postoperatively. The aim of this study was to compare and optimise the drill-in-behaviour of the EcoFit cup in a biomechanical investigation. EcoFit cup sizes 46, 50 and 54 mm were compared with the SC cup (Aesculap) size 50 mm. Foam blocks (Sawbones) of density 0.16 g/ml (pcf 10), 0.32 g/ml (pcf 20) and 0.48 g/ml (pcf 30) were used. After standardised placement using a universal testing system (n = 8 per group), the primary stability, the overhang of the cups and the drill-in behaviour were measured. Overreamings of 1 and 2 mm were performed (pcf 20, n = 8) for the EcoFit cup size 50 and the primary stability as well as the overhang measurements were examined. Measurements of the cup diameter, thread depth and thread pitch were performed on three-dimensional (3D) images of the cup size 50 mm. The drill-in behaviour was different between the EcoFit and the SC cups. Even with maximum torque, the EcoFit cup could not be positioned as deep as the SC cup in standard reaming conditions (overhang of 1.1 ± 0.4 mm for the EcoFit size 50 in pcf 20 and of -0.01 ± 0.2 mm for the SC cup). The primary stability was lower for the EcoFit cup in comparison to the SC cup (128.8 ± 3.2 Nm vs. 138.6 ± 9.1 Nm, p = 0.0291). With overreaming to 51 mm, a deeper positioning of the EcoFit was possible (overlap of -0.3 ± 0.1, comparable to the SC cup). The overreaming of the cavity also led to a significantly higher primary stability of 143.4 ± 3.7 Nm (p < 0.001) comparable to the unaltered condition (128.8 ± 3.2 Nm). Overreaming to 52 mm had no further advantage in terms of primary stability or overhang. The geometric measurements showed significant differences as well. The previously clinically observed difficulties in inserting the cup were confirmed by this study. By overreaming to 51 mm, the drill-in behaviour, the primary stability and the measured overhang were comparable to the reference cup. The obtained results suggest that the extension of the acetabulum cavity to 51 mm while using the implantcast EcoFit size 50 should be implemented in clinical applications.

Download full-text PDF

Source
http://dx.doi.org/10.1515/bmt-2019-0016DOI Listing

Publication Analysis

Top Keywords

ecofit cup
20
primary stability
16
cup
14
drill-in behaviour
12
g/ml pcf
12
ecofit
9
threaded cup
8
cup size
8
pcf
5
optimisation drill-in
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!