Purpose: The decay characteristics of radionuclides in PET studies can impact image reconstruction. Sc has been the topic of recent research due to potential theranostic applications and is a promising radiometal for PET imaging. In this study, the reconstructed images from phantom measurements with scandium in a small-animal PET scanner are compared with F and two prominent radiometals: Cu and Ga METHODS: Three phantoms filled with F, C, Ga, and Sc were imaged in the Siemens Inveon PET scanner. The NEMA image quality phantom was used to determine the recovery coefficients (RCs), spill-over ratios (SORs), and noise (%SD) under typical pre-clinical imaging conditions. Image contrast was determined using a Derenzo phantom, while the coincidence characteristics were investigated using an NEC phantom. Three reconstruction algorithms were used, namely filtered back projection (FBP), ordered subset expectation maximization (OSEM), and maximum a-posteriori (MAP).
Results: Image quality parameters were measured for F, Cu, Ga, and Sc respectively; using FBP, the %SD are 5.65, 5.88, 7.28, and 7.70; the RCs for the 5-mm rod are 0.849, 1.01, 0.615, and 0.825; the SORs in water are 0.0473, 0.0595, 0.141, 0.0923; and the SORs in air are 0.0589, 0.0484, 0.0525, and 0.0509. The contrast measured in the 2.5-mm rods are 0.674, 0.637, 0.196, and 0.347. The NEC rate with Sc increased at a slower rate than F and Ga as a function of activity in the field of view.
Conclusion: Sc demonstrates intermediate behavior relative to F and Ga with regard to RC and contrast measurements. It is a promising radionuclide for preclinical imaging.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6908536 | PMC |
http://dx.doi.org/10.1186/s40658-019-0260-0 | DOI Listing |
Cancer Biother Radiopharm
January 2025
Department of Nuclear Medicine, Hamidiye Faculty of Medicine, Prof. Dr. Cemil Taşcıoğlu City Hospital, University of Health Sciences, İstanbul, Türkiye.
This retrospective multicenter study investigated the biodistribution of Fluorodeoxyglucose (F-FDG) in the positron emission tomography (PET)/computed tomography (CT) in digital PET/CT (dPET) compared to analog PET/CT (aPET), focusing differences in physiological uptake in reference and small structures across various scanner models. One hundred thirty patients with similar preimaging conditions underwent both dPET and aPET imaging within 6 months. Visual evaluations and paired comparative analyses of semiquantitative parameters were performed for small and reference structures.
View Article and Find Full Text PDFJ Alzheimers Dis
January 2025
Health and Biosecurity, Commonwealth Scientific and Industrial Research Organisation, Brisbane, QLD, Australia.
Background: The introduction of therapeutics for Alzheimer's disease has led to increased interest in precisely quantifying amyloid-β (Aβ) burden for diagnosis, treatment monitoring, and further clinical research. Recent positron emission tomography (PET) hardware innovations including digital detectors have led to superior resolution and sensitivity, improving quantitative accuracy. However, the effect of PET scanner on Centiloid remains relatively unexplored and is assumed to be minimized by harmonizing PET resolutions.
View Article and Find Full Text PDFAcad Radiol
January 2025
Department of Radiology, Washington University School of Medicine, 510 S. Kingshighway Blvd, St. Louis, MO 63110 (S.I., M.A.T., M.I., C.S., R.L., A.H., R.L.W., T.J.F.). Electronic address:
Rationale And Objective: Conventional positron emission tomography (PET) respiratory gating utilizes a fraction of acquired PET counts (i.e., optimal gate [OG]), whereas elastic motion correction with deblurring (EMCD) utilizes all PET counts to reconstruct motion-corrected images without increasing image noise.
View Article and Find Full Text PDFCancers (Basel)
January 2025
Keimyung University School of Medicine, Daegu 42601, Republic of Korea.
Background/objectives: Accurate diagnosis is essential to avoid unnecessary procedures for thyroid incidentalomas (TIs). Advances in radiomics and machine learning applied to medical imaging offer promise for assessing thyroid nodules. This study utilized radiomics analysis on F-18 FDG PET/CT to improve preoperative differential diagnosis of TIs.
View Article and Find Full Text PDFMed Phys
January 2025
Faculty of Physics, Astronomy and Applied Computer Science, Jagiellonian University, Kraków, Poland.
Background: Total-body (TB) Positron Emission Tomography (PET) is one of the most promising medical diagnostics modalities, opening new perspectives for personalized medicine, low-dose imaging, multi-organ dynamic imaging or kinetic modeling. The high sensitivity provided by total-body technology can be advantageous for novel tomography methods like positronium imaging, demanding the registration of triple coincidences. Currently, state-of-the-art PET scanners use inorganic scintillators.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!