Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Introduction: A temporary myopic shift is a well-recognized complication of hyperbaric oxygen treatment (HBOT). Oxidation of proteins in the crystalline lens is the likely cause. Direct exposure of the eye to hyperbaric oxygen may exacerbate the effect. Our aim was to measure the magnitude of the myopic shift over a course of HBOT when using two different methods of oxygen delivery.
Methods: We conducted a randomised trial of oxygen delivery via hood versus oronasal mask during a course of 20 and 30 HBOT sessions. Subjective refraction was performed at baseline and after 20 and 30 sessions. We repeated these measurements at four and 12 weeks after completion of the course in those available for assessment.
Results: We enrolled 120 patients (mean age 57.6 (SD 11.2) years; 81% male). The myopic shift was significantly greater after both 20 and 30 sessions in those patients using the hood. At 20 treatments: refractory change was -0.92 D with hood versus -0.52 D with mask, difference 0.40 D (95% CI 0.22 to 0.57, P < 0.0001); at 30 treatments: -1.25 D with hood versus -0.63 with mask, difference 0.62 D (95% CI 0.39 to 0.84, P < 0.0001). Recovery was slower and less complete in the hood group at both four and 12 weeks.
Conclusions: Myopic shift is common following HBOT and more pronounced using a hood system than an oronasal mask. Recovery may be slower and less complete using a hood. Our data support the use of an oronasal mask in an air environment when possible.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7039782 | PMC |
http://dx.doi.org/10.28920/dhm49.4.245-252 | DOI Listing |