Beam commissioning of the first compact proton therapy system with spot scanning and dynamic field collimation.

Br J Radiol

Department of Radiotherapy (MAASTRO Proton Therapy), GROW-School for Oncology and Developmental Biology, Maastricht University Medical Centre, Maastricht, The Netherlands.

Published: March 2020

Objectives: To describe the measurements and to present the results of the beam commissioning and the beam model validation of a compact, gantry-mounted, spot scanning proton accelerator system with dynamic layer-by-layer field collimation.

Methods: We performed measurements of depth dose distributions in water, spot and scanned field size in air at different positions from the isocenter plane, spot position over the 20 × 20 cm scanned area, beam monitor calibration in terms of absorbed dose to water and specific field collimation measurements at different gantry angles to commission the system. To validate the beam model in the treatment planning system (TPS), we measured spot profiles in water at different depths, absolute dose in water of single energy layers of different field sizes and inversely optimised spread-out Bragg peaks (SOBP) under normal and oblique beam incidence, field size and penumbra in water of SOBPs, and patient treatment specific quality assurance in homogeneous and heterogeneous phantoms.

Results: Energy range, spot size, spot position and dose output were consistent at all gantry angles with 0.3 mm, 0.4 mm, 0.6 mm and 0.5% maximum deviations, respectively. Uncollimated spot size (one sigma) in air with an air-gap of 10 cm ranged from 4.1 to 16.4 mm covering a range from 32.2 to 1.9 cm in water, respectively. Absolute dose measurements were within 3% when comparing TPS and experimental data. Gamma pass rates >98% and >96% at 3%/3 mm were obtained when performing 2D dose measurements in homogeneous and in heterogeneous media, respectively. Leaf position was within ±1 mm at all gantry angles and nozzle positions.

Conclusions: Beam characterisation and machine commissioning results, and the exhaustive end-to-end tests performed to assess the proper functionality of the system, confirm that it is safe and accurate to treat patients.

Advances In Knowledge: This is the first paper addressing the beam commissioning and the beam validation of a compact, gantry-mounted, pencil beam scanning proton accelerator system with dynamic layer-by-layer multileaf collimation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7066978PMC
http://dx.doi.org/10.1259/bjr.20190598DOI Listing

Publication Analysis

Top Keywords

beam commissioning
12
gantry angles
12
beam
10
spot
8
spot scanning
8
field collimation
8
commissioning beam
8
beam model
8
validation compact
8
compact gantry-mounted
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!