Signal reversal in Kelvin-probe force microscopy.

Rev Sci Instrum

Automation and Control Institute (ACIN), TU Wien, Gusshausstrasse 27-29, A-1040 Vienna, Austria.

Published: November 2019

Kelvin-probe force microscopy is a measurement mode of atomic force microscopy, which is used to quantitatively map the electrical surface potential of a sample. Inadequate hardware and electronic design can lead to signal cross talk and, in consequence, false results. Here, we show that certain cross talk artifacts not only do manifest themselves in additional noise, reduced resolution, or an offset of the measured surface potential but can also lead to an inverted signal scale and, crucially, cannot be diagnosed with a known reference signal. We show experimental data on an electrically homogeneous sample, describe a method to detect the artifact, and propose simple remedies, which should be well within the reach of most research and industrial laboratories.

Download full-text PDF

Source
http://dx.doi.org/10.1063/1.5118357DOI Listing