Introduction: Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) has been introduced for bacterial identification. The ASTA MicroIDSys system (ASTA, Suwon, Korea) is a new MALDI-TOF MS system developed for species identification of microorganisms. We evaluated the performance of MicroIDSys against clinical isolates of anaerobic bacteria.

Material And Methods: A total of 370 non-duplicated clinical isolates of anaerobic bacteria were tested in this study. Bacterial identification with MicroIDSys was performed with a direct smear method, and measured spectra were analyzed using respective software. The results of MicroIDSys were compared with the results of Bruker Biotyper and 16S rRNA sequencing.

Results: The overall agreement rates for the 370 clinical isolates (34 genera and 99 species) were 95.4% (353/370) at the genus level and 91.6% (n = 340) at the species level. Only 17 isolates were incorrectly identified at the genus level: five misidentifications and 12 unidentifications. The MicroIDSys system exhibited excellent performance in the identification of clinically relevant bacterial species. Most of the Bacteroides isolates (98.0%, 99/101) and all of the Clostridium difficile (100%, n = 11), Clostridium perfringens (100%, n = 10), Finegoldia magna (100%, n = 11), and Parvimonas micra (100%, n = 10) isolates were correctly identified at the species level.

Conclusion: The MicroIDSys system proved useful in the identification of anaerobic bacteria, especially clinically relevant species. This system could be of use in clinical microbiology laboratories as a primary tool for identifying anaerobic bacteria.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.anaerobe.2019.102131DOI Listing

Publication Analysis

Top Keywords

microidsys system
16
clinical isolates
16
anaerobic bacteria
16
bacterial identification
12
isolates anaerobic
12
matrix-assisted laser
8
laser desorption/ionization
8
desorption/ionization time-of-flight
8
time-of-flight mass
8
mass spectrometry
8

Similar Publications

Clinical Evaluation of VITEK MS PRIME with PICKME Pen for Bacteria and Yeasts, and RUO Database for Filamentous Fungi.

Microorganisms

May 2024

Department of Laboratory Medicine, International St. Mary's Hospital, College of Medicine, Catholic Kwandong University, Incheon 22711, Republic of Korea.

The VITEK MS PRIME (bioMérieux, Marcy-l'Étoile, France), a newly developed matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) system, alongside the VITEK PICKME pen (PICKME), offers easy sample preparation for bacteria and yeasts. The VITEK MS PRIME also offers two software platforms for filamentous fungi: the IVD database and the RUO database. Our study evaluated its identification agreement on 320 clinical isolates of bacteria and yeasts, comparing PICKME and traditional wooden toothpick sampling techniques against MicroIDSys Elite (ASTA) results.

View Article and Find Full Text PDF

Purpose: species can cause infections, and associations with cancer are being increasingly reported. As their clinical significance differs, accurate identification of individual species is important. However, matrix-assisted laser desorption/ionization-time of flight mass spectrometry has not been found to be effective in identifying species in previous studies.

View Article and Find Full Text PDF

Identification of the rare yeast Cutaneotrichosporon (Trichosporon) debeurmannianum from diabetic foot infection.

J Clin Lab Anal

December 2022

Department of Laboratory Medicine, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, South Korea.

Background: Diabetic foot infection is the most common complications of diabetes mellitus. Although most of the diabetic foot infections has been known to be caused by aerobic and anaerobic bacteria, mycotic diabetic foot infection caused by Candida species has also been reported recently. Here, we present the first case of diabetic foot infection caused by Cutaneotrichosporon debeurmannianum (previously known as Trichosporon debeurmannianum).

View Article and Find Full Text PDF

Filamentous fungi are a major cause of life-threatening infections in immunocompromised patients; thus, rapid and accurate identification is critical. Filamentous fungal identification by matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) has been demonstrated with high sensitivity and reproducibility; however, its wider application has been limited in clinical laboratories because of practical challenges such as database availability or lack of standardization. In this study, we compared the performance of the Bruker Biotyper, ASTA MicroIDSys, and Vitek MS for 84 clinical filamentous fungal isolates.

View Article and Find Full Text PDF

This study compared the accuracy of a new MALDI-TOF mass spectrometry system, ASTA MicroIDSys system, with that of MALDI Biotyper system for the identification of reference and clinical bacterial and yeast strains. The identification accuracy of the 2 systems was compared using a total of 406 strains comprising 142 aerobic and 180 anaerobic bacterial strains and 84 yeast strains. The genus and species identification rates were 98.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!