Background: A new point-of-care (POC) HIV virus load technology has been recently developed and designed to be utilized in decentralized settings. Alere Technologies GmbH*, Germany, developed the mPIMA HIV-1/2 VL plasma test which uses real time PCR technology with 50 μl and a turnaround time of one hour.
Objective: Analyze the performance of mPIMA to detect and quantify HIV-1 and HIV-2 and compare with Abbott M2000 assay fooling patients HIV-1 failing ARV therapy.
Study Design: In this study we evaluate the mPIMA HIV-1/2 VL plasma test using 413 specimens from 270 patients failing ARV therapy, and compared its performance with Abbott RealTime HIV-1 Viral Load assay on the m2000 system. In addition, were determined VL in plasma specimens obtained from HIV-2 infected patients.
Results: The results strongly indicate that mPIMA HIV-1/2 VL plasma test can determine HIV-1 with concordance of 88.9 % (95 % CI 85.4-91.7) the reference test when 1000 HIV-1 VL threshold was used as WHO cutoff to identify therapy failure. The overall correlation between HIV-1 VL was 0928 (Pearson correlation coefficient of Linear regression) and the Bland-Altman showed a mean difference of -0.20 Log cp/mL between the two technology. mPIMA HIV-1/2 VL plasma test was also able to measure HIV-2 viral load in 16 specimens from Guinea-Bissau HIV-1/HIV-2 positive samples.
Conclusions: These data support the use of mPIMA HIV-1/2 VL plasma test to follow up patients and select patients failing ART, guiding immediate clinical decisions such as adherence counseling or ART regimen switch during the patient consultation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcv.2019.104212 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!