Aims: We sought to investigate the thrombogenicity of different DES and BMS in an in vitro system of stent perfusion.

Material And Methods: The experimental model consisted of a peristaltic pump connected to 4 parallel silicone tubes in which different stents were deployed. Blood was drawn from healthy volunteers and the amount of stent surfaced-induced thrombus deposition was determined using I-fibrinogen.

Results: Compared to Resolute, Biomatrix and Vision, Xience was associated with the lowest amount of stent surface-induced thrombus formation, with a significant difference compared to Vision (I-fibrinogen median value deposition [IQ range]: 50 ng [25-98] versus 560 ng [320-1520], respectively, p < 0.05), but not to other DES. In the second set of experiments Fluoropolymer-coated BMS not eluting drug was associated with a significant 3-fold reduction in I-fibrinogen deposition (245 ng [80-300]) compared to Vision (625 ng [320-760], p < 0.05), but a 7-fold increase compared to Xience (35 ng [20-60], p < 0.05). Finally Xience was associated with a significantly greater absorption of albumin compared to BMS.

Conclusions: In an in vitro system of stent perfusion, Xience was associated with the lowest amount of stent surface-induced thrombus formation compared with Resolute, Biomatrix and Vision, with a noted synergistic effect between the fluoropolymer and the drug.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.thromres.2019.11.016DOI Listing

Publication Analysis

Top Keywords

amount stent
8
vitro thrombogenicity
4
thrombogenicity drug-eluting
4
drug-eluting bare
4
bare metal
4
metal stents
4
stents aims
4
aims sought
4
sought investigate
4
investigate thrombogenicity
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!