Purpose: To describe and validate the dose calculation algorithm of an independent second-dose check software for spot scanning proton delivery systems with full width at half maximum between 5 and 14 mm and with a negligible spray component.
Methods: The analytical dose engine of our independent second-dose check software employs an altered pencil beam algorithm with 3 lateral Gaussian components. It was commissioned using Geant4 and validated by comparison to point dose measurements at several depths within spread-out Bragg peaks of varying ranges, modulations, and field sizes. Water equivalent distance was used to compensate for inhomogeneous geometry. Twelve patients representing different disease sites were selected for validation. Dose calculation results in water were compared to a fast Monte Carlo code and ionization chamber array measurements using dose planes and dose profiles as well as 2-dimensional-3-dimensional and 3-dimensional-3-dimensional γ-index analysis. Results in patient geometry were compared to Monte Carlo simulation using dose-volume histogram indices, 3-dimensional-3-dimensional γ-index analysis, and inpatient dose profiles.
Results: Dose engine model parameters were tuned to achieve 1.5% agreement with measured point doses. The in-water γ-index passing rates for the 12 patients using 3%/2 mm criteria were 99.5% ± 0.5% compared to Monte Carlo. The average inpatient γ-index analysis passing rate compared to Monte Carlo was 95.8% ± 2.9%. The average difference in mean dose to the clinical target volume between the dose engine and Monte Carlo was -0.4% ± 1.0%. For a typical plan, dose calculation time was 2 minutes on an inexpensive workstation.
Conclusions: Following our commissioning process, the analytical dose engine was validated for all treatment sites except for the lung or for calculating dose-volume histogram indices involving point doses or critical structures immediately distal to target volumes. Monte Carlo simulations are recommended for these scenarios.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6876166 | PMC |
http://dx.doi.org/10.1177/1533033819887182 | DOI Listing |
Pharm Stat
January 2025
Department of Anesthesia, University of Iowa, Iowa City, Iowa, USA.
With contemporary anesthetic drugs, the efficacy of general anesthesia is assured. Health-economic and clinical objectives are related to reductions in the variability in dosing, variability in recovery, etc. Consequently, meta-analyses for anesthesiology research would benefit from quantification of ratios of standard deviations of log-normally distributed variables (e.
View Article and Find Full Text PDFJ Biomed Opt
January 2025
CIFICEN (UNCPBA - CICPBA - CONICET), Tandil, Argentina.
Significance: In the last years, time-resolved near-infrared spectroscopy (TD-NIRS) has gained increasing interest as a tool for studying tissue spectroscopy with commercial devices. Although it provides much more information than its continuous wave counterpart, accurate models interpreting the measured raw data in real time are still lacking.
Aim: We introduce an analytical model that can be integrated and used in TD-NIRS data processing software and toolkits in real time.
Monte Carlo Methods Appl
December 2024
Computer Languages and Systems Software Group, Lawrence Berkeley National Laboratory, Berkeley, California, USA.
Correlation coefficients and linear regression values computed from group averages can differ from correlation coefficients and linear regression values computed using individual scores. This observation known as the ecological fallacy often assumes that all the individual scores are available from a population. In many situations, one must use a sample from the larger population.
View Article and Find Full Text PDFHeliyon
January 2025
Environmental Energy Technologies Laboratory (EETL), Department of Physics, University of Yaounde I, P.O Box 812 Yaounde, Cameroon.
This article analyzes and compares three methodologies for identifying suitable regions for solar hydrogen production using photovoltaic panels: AHP (Analytic Hierarchy Process), FAHP (Fuzzy Analytic Hierarchy Process), and MC-FAHP (Monte Carlo FAHP), integrated with GIS (Geographic Information Systems). The study employs ten criteria across technical (Global Horizontal Irradiance, temperature, slope, elevation, orientation), economic (distance from transportation and electrical networks), and social (population density, proximity to residential areas) factors. Environmental and exclusion criteria define restrictive zones.
View Article and Find Full Text PDFNMR Biomed
March 2025
Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
The purpose of this study was to measure T and T relaxation times of NAD proton resonances in the downfield H MRS spectrum in human brain at 7 T in vivo and to assess the propagation of relaxation time uncertainty in NAD quantification. Downfield spectra from eight healthy volunteers were acquired at multiple echo times to measure T relaxation times, and saturation recovery data were acquired to measure T relaxation times. The downfield acquisition used a spectrally selective 90° sinc pulse for excitation centered at 9.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!