Objectives: We aim to develop and rigorously evaluate an image-based deconvolution method to jointly compensate respiratory motion and partial volume effects (PVEs) for quantitative oncologic PET imaging, including studying the impact of various reconstruction algorithms on quantification performance.
Procedures: An image-based deconvolution method that incorporated wavelet-based denoising within the Lucy-Richardson algorithm was implemented and assessed. The method was evaluated using phantom studies with signal-to-background ratios (SBR) of 4 and 8, and clinical data of 10 patients with 42 lung lesions ≤30 mm in diameter. In each study, PET images were reconstructed using four different algorithms: OSEM-basic, PSF, TOF, and TOFPSF. The performance was quantified using contrast recovery (CR), coefficient of variation (COV) and contrast-to-noise-ratio (CNR) metrics. Further, in each study, variabilities arising due to the four different reconstruction algorithms were assessed.
Results: In phantom studies, incorporation of wavelet-based denoising improved COV in all cases. Processing images using proposed method yielded significantly higher CR and CNR particularly in small spheres, for all reconstruction algorithms and all SBRs (P < 0.05). In patient studies, processing images using the proposed method yielded significantly higher CR and CNR (P < 0.05). The choice of the reconstruction algorithm impacted quantification performance for changes in motion amplitude, tumor size and SBRs.
Conclusions: Our results provide strong evidence that the proposed joint-compensation method can yield improved PET quantification. The choice of the reconstruction algorithm led to changes in quantitative accuracy, emphasizing the need to carefully select the right combination of reconstruction-image-based compensation methods.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.ejmp.2019.10.031 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!