Background: The optimal programmed intermittent epidural bolus regimen for labor analgesia remains unknown. Some studies indicate that better drug spread in the epidural space results from greater injection pressure; however, there is a lack of data regarding the maximum pressure generated by epidural bolus injection using different catheters and flow rates.
Methods: We evaluated the flow and pressure characteristics of 11 commonly used epidural catheters combined with 3 different infusion pumps that deliver epidural infusions according to the programmed intermittent epidural bolus regimen. Pressure changes were measured over time at flow rates of 100, 250, and 400 mL·hour and with a bolus volume of 10 mL. To account for repeated measures, linear mixed models were used. Features were selected with a backward stepwise procedure continued until only statistically significant variables were left in the model.
Results: We performed 660 measurements. The mean maximal pressure generated during bolus injection ranged from 86 to 863 mm Hg for different flow rates and catheter designs. The interaction between flow rate and catheter gauge resulted in 1.31, 1.65, and 2.00 mm Hg of pressure increase for 18G, 19G, and 20G catheters, respectively, per 1 mL·hour of increased flow rate (P< .001). Analyses including wire-reinforced catheters revealed a 1.16, 1.76, and 2.36 mm Hg pressure increase for 18G, 19G, and 20G catheters, respectively, per 1 mL·hour of increased flow rate (P< .001). In some cases, it triggered the occlusion pump alarm.
Conclusions: Significant differences were observed in the in vitro maximum pressure value among the various catheter and flow rate combinations with a higher pressure value for wire-reinforced catheters used in the study. The optimal flow rate and epidural catheter combination may allow for delivery of the bolus with high flow rate without triggering the occlusion alarm.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1213/ANE.0000000000003650 | DOI Listing |
Heliyon
January 2025
Institute of Mathematics, Henan Academy of Sciences, Zhengzhou, 450046, China.
This study examines the behavior of the Casson nanofluid bioconvection flow around a spinning disc under various influences, including gyrotactic microorganisms, multiple slips, and thermal radiation. Notably, it accounts for the reversible nature of the flow and incorporates the esterification process. The aim of this study is to investigate the influence of reversible chemical reactions on the flow behavior of a Casson nanofluid in the presence of bioconvective microorganisms over a spinning disc.
View Article and Find Full Text PDFInt J Exerc Sci
December 2024
Exercise Science, Florida Southern College, Lakeland, FL, USA.
The purpose of the study was to compare heart rate (HR), systolic blood pressure (SBP), and diastolic blood pressure (DBP) following high load resistance exercise (HLRE) and blood flow restriction exercise (BFRE) with a knee wrap (kBFRE) and pneumatic cuff (pBFRE). Eleven men (N = 9) and women (N = 2) participated. HR, SBP, and DBP were collected at Rest, immediately post exercise (IP), 10-, 30-, and 45-minutes post exercise.
View Article and Find Full Text PDFJ Soc Cardiovasc Angiogr Interv
December 2024
The University of Toledo, Toledo, Ohio.
Background: Atherosclerotic renal artery stenosis (ARAS) may provoke hypertension and/or impaired kidney function. Some patients develop uncontrolled hypertension and deteriorating kidney function despite optimal medical therapy. In these patients, endovascular treatment is an important therapeutic option.
View Article and Find Full Text PDFAnticancer Agents Med Chem
January 2025
Department of Biology, Faculty of Science, Atatürk University, Erzurum, Turkey.
Introduction/objective: Several nutraceuticals, food, and cosmetic products can be developed using royal jelly. It is known for its potential health benefits, including its ability to boost the immune system and reduce inflammation. It is rich in vitamins, minerals, and antioxidants, which can improve general health.
View Article and Find Full Text PDFJ Sep Sci
January 2025
R & D Laboratory, TASNEE Technology and Innovation Centre, Al-Jubail Industrial City, Saudi Arabia.
In this study, a commercially available polypropylene homopolymer (H-PP) was blended with blow molding polyethylene (PE) grade via melt mixing using a compounding machine. The resulting blends were subjected to high-temperature size exclusion chromatography (SEC) analysis, coupled with infrared-5 (IR-5), viscometer (VISCO), and multi-angle laser light scattering (MALS) detectors. The molecular weight (MW) and MW distributions were investigated using SEC, and the exact blend compositions were evaluated using C nuclear magnetic resonance.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!