Purpose: To evaluate the use of the absorbed depth-dose as a surrogate of the half-value layer in the calibration of a high-dose-rate electronic brachytherapy (eBT) equipment. The effect of the manufacturing tolerances and the absorbed depth-dose measurement uncertainties in the calibration process are also addressed.

Methods: The eBT system Esteya (Elekta Brachytherapy, Veenendaal, The Netherlands) has been chosen as a proof-of-concept to illustrate the feasibility of the proposed method, using its 10 mm diameter applicator. Two calibration protocols recommended by the AAPM (TG-61) and the IAEA (TRS-398) for low-energy photon beams were evaluated. The required Monte Carlo (MC) simulations were carried out using PENELOPE2014. Several MC simulations were performed modifying the flattening filter thickness and the x-ray tube potential, generating one absorbed depth-dose curve and a complete set of parameters required in the beam calibration (i.e., HVL, backscatter factor (B ), and mass energy-absorption coefficient ratios (µ /ρ) ), for each configuration. Fits between each parameter and some absorbed dose-ratios calculated from the absorbed depth-dose curves were established. The effect of the manufacturing tolerances and the absorbed dose-ratio uncertainties over the calibration process were evaluated by propagating their values over the fitting function, comparing the overall calibration uncertainties against reference values. We proposed four scenarios of uncertainty (from 0% to 10%) in the dose-ratio determination to evaluate its effect in the calibration process.

Results: The manufacturing tolerance of the flattening filter (±0.035 mm) produces a change of 1.4% in the calculated HVL and a negligible effect over the B , (µ /ρ) , and the overall calibration uncertainty. A potential variation of 14% of the electron energies due to manufacturing tolerances in the x-ray tube (69.5 ± ~10 keV) generates a variation of 10% in the HVL. However, this change has a negligible effect over the B and (µ /ρ) , adding 0.1% to the overall calibration uncertainty. The fitting functions reproduce the data with an uncertainty (k = 2) below 1%, 0.5%, and 0.4% for the HVL, B , and (µ /ρ) , respectively. The four studied absorbed dose-ratio uncertainty scenarios add, in the worst-case scenario, 0.2% to the overall uncertainty of the calibration process.

Conclusions: This work shows the feasibility of using the absorbed depth-dose curve in the calibration of an eBT system with minimal loss of precision.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.13920DOI Listing

Publication Analysis

Top Keywords

absorbed depth-dose
24
calibration
12
manufacturing tolerances
12
absorbed
9
beam calibration
8
tolerances absorbed
8
uncertainties calibration
8
calibration process
8
ebt system
8
flattening filter
8

Similar Publications

Commissioning of a commercial treatment planning system for scanned carbon-ion radiotherapy.

J Appl Clin Med Phys

November 2024

Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Shanghai Key Laboratory of Radiation Oncology, Shanghai Engineering Research Center of Proton and Heavy Ion Radiation Therapy, Shanghai, China.

Purpose: To commission the RayStation (RS) TPS (treatment planning system) for scanned CIRT (carbon-ion radiotherapy) utilizing pencil beam algorithms (PBv4.2).

Methods: The beam model commissioning entailed employing 1D single beams and 2D monoenergetic fields to validate spot profiles with films, assess beam range using Peakfinder measurements, and evaluate fragment spectra through dose-averaged linear energy transfer (LETd) calculations.

View Article and Find Full Text PDF

Background And Purpose: Radiotherapy using Very High Energy Electrons (VHEE) has the potential to reduce dose to organs at risk compared to photons. This article therefore reviews treatment planning for VHEE, to clarify the potential benefit of the modality.

Materials And Methods: Articles on VHEE were identified and those which focused on treatment planning were manually selected, particularly those which contained results on patient datasets.

View Article and Find Full Text PDF

The role of volume averaging effects, beam hardening, and phantom scatter in dosimetry of grid therapy.

Phys Med Biol

November 2024

Department of Radiation Oncology, Northwestern Memorial Hospital, Northwestern University Feinberg School of Medicine, Chicago, IL 60611, United States of America.

. Current reference dosimetry methods for spatially fractionated radiation therapy (SFRT) assume a negligible beam quality change, perturbation, or volume-averaging correction factor. Therefore, the aim of this work was to investigate the impact of the grid collimators on the dosimetric characteristics of a 6 MV photon beam.

View Article and Find Full Text PDF

FLASH radiotherapy using high-energy X-rays: Current status of PARTER platform in FLASH research.

Radiother Oncol

October 2023

Department of Radiotherapy Physics & Technology, West China Hospital, Sichuan University, Chengdu 610041, China. Electronic address:

Purpose: Recent studies indicated that ultrahigh dose rate (FLASH) radiation can reduce damage to normal tissue while maintaining anti-tumour activity compared to conventional dose rate (CONV) radiation. This paper provides a comprehensive description of the current status of the Platform for Advanced Radiotherapy Research (PARTER), which serves as the first experimental FLASH platform utilizing megavoltage X-rays and has facilitated numerous experiments.

Methods And Materials: PARTER was established in 2019 based on a superconducting linac to support experimental FLASH studies using megavoltage X-rays.

View Article and Find Full Text PDF

A 3D range-modulator (RM), optimized for a single energy and a specific target shape, is a promising and viable solution for the ultra-fast dose delivery in particle therapy. The aim of this work was to investigate the impact of potential beam and modulator misalignments on the dose distribution. Moreover, the FLUKA Monte Carlo model, capable of simulating 3D RMs, was adjusted and validated for the 250 MeV single-energy proton irradiation from a Varian ProBeam system.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!