Background: Little data are available about radiation exposure during cardiac electrical device implantation, and no dose reference levels have been published. This multicenter, prospective, observational study assesses patient and staff radiation exposure during cardiac device implantations, and aims at defining dose reference levels.

Methods: Patient demographic, procedural, and radiation data were obtained for 657 procedures from nine institutions. Physician and staff exposure were measured using real-time dosimeters worn beneath and above lead apron. Statistical analysis included fluoroscopy time (FT), dose-area product (DAP), and DAP adjusted for FT and body mass index.

Results: Pacemakers and cardioverter defibrillators were implanted in 481 and 176 patients, respectively. Of these, 152 were treated with cardiac resynchronization therapy (CRT). Median FTs were 837s (interquartile range [IQR]: 480-1323), 117s (IQR: 69-209), and 101s (IQR: 58-162), and median DAPs were 1410 (IQR: 807-2601), 150 (IQR: 72-338), and 129 (IQR: 72-332) cGy.cm² for biventricular, dual chamber, and ventricular device implantation, respectively. Dose reference levels correspond to the third quartile values. During CRT, higher exposure was observed with four X-ray systems than with the two newer and customizable ones (adjusted DAP of 0.90 [IQR: 0.26-1.01] and 0.29 [IQR: 0.23-0.39], respectively; P < .001).

Conclusion: Based on real-life measurements, this multicenter registry provides dose reference levels and may help centers assess radiation exposure. Although biventricular device implantation was responsible for the highest radiation exposure, FT was meaningfully shortened compared to previously reported values. For a same FT, the use of new generators and custom settings has significantly reduced DAP.

Download full-text PDF

Source
http://dx.doi.org/10.1111/pace.13842DOI Listing

Publication Analysis

Top Keywords

radiation exposure
12
exposure cardiac
12
device implantation
12
dose reference
12
cardiac device
8
implantation dose
8
reference levels
8
iqr
5
radiation
4
cardiac
4

Similar Publications

Cancer remains a leading cause of mortality worldwide, requiring physicians to understand multidisciplinary treatments. This study assessed the impact of a clinical rotation in a cancer center on medical students' knowledge of cancer treatments from a multidisciplinary perspective. A traditional single-department rotation was compared to a multidisciplinary rotation to determine whether broader exposure enhances knowledge and prepares students for multidisciplinary care.

View Article and Find Full Text PDF

Effect of terahertz radiation on cells and cellular structures.

Front Optoelectron

January 2025

Institute of Physics, Saratov State University, Saratov, 410012, Russia.

The paper presents the results of modern research on the effects of electromagnetic terahertz radiation in the frequency range 0.5-100 THz at different levels of power density and exposure time on the viability of normal and cancer cells. As an accompanying tool for monitoring the effect of radiation on biological cells and tissues, spectroscopic research methods in the terahertz frequency range are described, and attention is focused on the possibility of using the spectra of interstitial water as a marker of pathological processes.

View Article and Find Full Text PDF

Objective: To explore the efficacy and safety of the direct inferior endplate approach in percutaneous endoscopic interlaminar discectomy (PEID) for the treatment of L5-S1 disc herniation.

Methods: This was a retrospective analysis of 116 patients with L5-S1 disc herniation treated with PEID; 74 patients underwent surgery via the direct inferior endplate approach (group A), and 42 patients underwent surgery via the indirect approach (group B). The number of intraoperative fluoroscopy exposures, establishment channel time, operation time, postoperative visual analogue scale (VAS) score, and Oswestry Disability Index (ODI) were compared between the 2 groups.

View Article and Find Full Text PDF

Objectives: In advanced stages of osteoradionecrosis, medication-related osteonecrosis of the jaw, and osteomyelitis, a resection of sections of the mandible may be unavoidable. The determination of adequate bony resection margins is a fundamental problem because bony resection margins cannot be secured intraoperatively. Single-photon emission computed tomography (SPECT-CT) is more accurate than conventional imaging techniques in detecting inflammatory jaw pathologies.

View Article and Find Full Text PDF

Keratin/chitosan film promotes wound healing in rats with combined radiation-wound injury.

J Mater Sci Mater Med

January 2025

Department of Nuclear Medicine, Chongqing University Cancer Hospital, No. 181 HanYu St, Shapingba District, Chongqing, 400030, PR China.

Human hair keratin, a natural protein derived from human hair, has emerged prominently in the field of wound repair, showcasing its unique regenerative capabilities and extensive application potential. However, it is a challenge for the keratin to efficiently therapy the impaired wound healing, such as combined radiation-wound injury. Here, we report a keratin/chitosan (KRT/CS) film for skin repair of chronic wounds in in rats with combined radiation-wound injury.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!