In a pulsatile flow model aortic Björk-Shiley Standard, Convex-Concave and Monostrut valves were investigated together with the Hall-Kaster (Medtronic-Hall), St Jude Medical and Starr-Edwards Silastic Ball valve using hot-film anemometry. Three-dimensional visualization of average systolic Reynolds normal stresses (RNS) reflected the design of the valves. Mean average RNS were used for comparison of the fluid dynamic performance along with Velocity Energy Ratio (VER100) and Turbulence Energy Ratio (TER) as a relative turbulence intensity for pulsatile flow. Mean average RNS ranged from 13.2 to 37.6 Nm-2 for all the valves with the highest levels for the Björk-Shiley Standard and Starr-Edwards Ball valve and lowest values for the St Jude Medical valve and with the Hall-Kaster (Medtronic-Hall), Björk-Shiley Convex-Concave and Monostrut valves in between.