Low-flow assessment of current ECMO/ECCOR rotary blood pumps and the potential effect on hemocompatibility.

Crit Care

Department of Pneumology and Critical Care Medicine, Cologne-Merheim Hospital, ARDS and ECMO Center, Kliniken der Stadt Köln gGmbH, Witten/Herdecke University Hospital, Ostmerheimer Strasse 200, 51109, Cologne, Germany.

Published: November 2019

Background: Extracorporeal carbon dioxide removal (ECCOR) uses an extracorporeal circuit to directly remove carbon dioxide from the blood either in lieu of mechanical ventilation or in combination with it. While the potential benefits of the technology are leading to increasing use, there are very real risks associated with it. Several studies demonstrated major bleeding and clotting complications, often associated with hemolysis and poorer outcomes in patients receiving ECCOR. A better understanding of the risks originating specifically from the rotary blood pump component of the circuit is urgently needed.

Methods: High-resolution computational fluid dynamics was used to calculate the hemodynamics and hemocompatibility of three current rotary blood pumps for various pump flow rates.

Results: The hydraulic efficiency dramatically decreases to 5-10% if operating at blood flow rates below 1 L/min, the pump internal flow recirculation rate increases 6-12-fold in these flow ranges, and adverse effects are increased due to multiple exposures to high shear stress. The deleterious consequences include a steep increase in hemolysis and destruction of platelets.

Conclusions: The role of blood pumps in contributing to adverse effects at the lower blood flow rates used during ECCOR is shown here to be significant. Current rotary blood pumps should be used with caution if operated at blood flow rates below 2 L/min, because of significant and high recirculation, shear stress, and hemolysis. There is a clear and urgent need to design dedicated blood pumps which are optimized for blood flow rates in the range of 0.5-1.5 L/min.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6836552PMC
http://dx.doi.org/10.1186/s13054-019-2622-3DOI Listing

Publication Analysis

Top Keywords

blood pumps
20
rotary blood
16
blood flow
16
flow rates
16
blood
11
carbon dioxide
8
current rotary
8
adverse effects
8
shear stress
8
flow
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!