Objective: To investigate whether a microelectromechanical system (MEMS) inertial sensor module is as accurate as fiber-optic gyroscopes when classifying subjects as normal for clinical stance and gait balance tasks.

Methods: Data of ten healthy subjects were recorded simultaneously with a fiber-optic gyroscope (FOG) system of SwayStar™ and a MEMS sensor system incorporated in the Valedo® system. Data from a sequence of clinical balance tasks with different angle and angular velocity ranges were assessed. Paired -tests were performed to determine significant differences between measurement systems. Cohen's kappa test was used to determine the classification of normal balance control between the two sensor systems when comparing the results to a reference database recorded with the FOG system. Potential cross-talk errors in roll and pitch angles when neglecting yaw axis rotations were evaluated by comparing 2D FOG and 3D MEMS recordings.

Results: Statistically significant (=0.05) differences were found in some balance tasks, for example, "walking eight tandem steps" and various angular measures ( < 0.03). However, these differences were within a few percent (<2.7%) of the reference values. Tasks with high dynamic velocity ranges showed significant differences (=0.002) between 2D FOG and 3D MEMS roll angles but no difference between 2D FOG and 2D MEMS roll angles. An almost perfect agreement could be obtained for both 2D FOG and 2D MEMS (=0.97) and 2D FOG and 3D MEMS measures (=0.87) when comparing measurements of all subjects and tasks.

Conclusion: MEMS motion sensors can be used for assessing balance during clinical stance and gait tasks. MEMS provides measurements comparable to values obtained with a highly accurate FOG. When assessing pitch and roll trunk sway measures without accounting for the effect of yaw, it is recommended to use angle and angular velocity measures for stance, and only angular velocity measures for gait because roll and pitch velocity measurements are not influenced by yaw rotations, and angle errors are low for stance.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6778920PMC
http://dx.doi.org/10.1155/2019/9816961DOI Listing

Publication Analysis

Top Keywords

inertial sensor
8
sensor module
8
fiber-optic gyroscope
8
clinical balance
8
fog system
8
balance tasks
8
balance
5
system
5
comparison low-cost
4
low-cost miniature
4

Similar Publications

Introduction: Gait analysis is a vital tool in the assessment of human movement and has been widely used in clinical settings to identify potential abnormalities in individuals. However, there is a lack of consensus on the normative values for gait metrics in large populations. The primary objective of this study is to establish a normative database of spatiotemporal gait metrics across various age groups, contributing to a broader understanding of human gait dynamics.

View Article and Find Full Text PDF

Action Recognition in Basketball with Inertial Measurement Unit-Supported Vest.

Sensors (Basel)

January 2025

Computer Engineering Department, Engineering Faculty, Aydın Adnan Menderes University, Aydın 09100, Türkiye.

In this study, an action recognition system was developed to identify fundamental basketball movements using a single Inertial Measurement Unit (IMU) sensor embedded in a wearable vest. This study aims to enhance basketball training by providing a high-performance, low-cost solution that minimizes discomfort for athletes. Data were collected from 21 collegiate basketball players, and movements such as dribbling, passing, shooting, layup, and standing still were recorded.

View Article and Find Full Text PDF

To enhance the positioning accuracy of autonomous underwater vehicles (AUVs), a new adaptive filtering algorithm (RHAUKF) is proposed. The most widely used filtering algorithm is the traditional Unscented Kalman Filter or the Adaptive Robust UKF (ARUKF). Excessive noise interference may cause a decrease in filtering accuracy and is highly likely to result in divergence by means of the traditional Unscented Kalman Filter, resulting in an increase in uncertainty factors during submersible mission execution.

View Article and Find Full Text PDF

In hospitals, timely interventions can prevent avoidable clinical deterioration. Early recognition of deterioration is vital to stopping further decline. Measuring the way patients position themselves in bed and change their positions may signal when further assessment is necessary.

View Article and Find Full Text PDF

According to South Korea's Ministry of Employment and Labor, approximately 25,000 construction workers suffered from various injuries between 2015 and 2019. Additionally, about 500 fatalities occur annually, and multiple studies are being conducted to prevent these accidents and quickly identify their occurrence to secure the golden time for the injured. Recently, AI-based video analysis systems for detecting safety accidents have been introduced.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!