CBCT-Based Image Guidance for Percutaneous Access: Electromagnetic Navigation Versus 3D Image Fusion with Fluoroscopy Versus Combination of Both Technologies-A Phantom Study.

Cardiovasc Intervent Radiol

Assistance Publique - Hôpitaux de Paris (AP-HP), Service d'Imagerie Médicale, CHU Henri Mondor, 51 Avenue du Maréchal de Lattre de Tassigny, 94010, Créteil, France.

Published: March 2020

Purpose: We set out to compare three types of three-dimensional CBCT-based imaging guidance modalities in a phantom study: image fusion with fluoroscopy (IF), electromagnetic navigation (EMN) and the association of both technologies (CEMNIF).

Materials And Methods: Four targets with a median diameter of 11 mm [first quartile (Q1): 10; third quartile (Q3): 12] with acute angle access (z-axis < 45°) and four targets of 10 mm [8-15] with large angle access (z-axis > 45°) were defined on an abdominal phantom (CIRS, Meditest, Tabuteau, France). Acute angle access targets were punctured using IF, EMN or CEMNIF and large angle access targets with EMN by four operators with various experiences. Efficacy (target reached), accuracy (distance between needle tip and target center), procedure time, radiation exposure and reproducibility were explored and compared.

Results: All targets were reached (100% efficacy) by all operators. For targets with acute angle access, procedure times (EMN: 265 s [236-360], IF: 292 s [260-345], CEMNIF: 320 s [240-333]) and accuracy (EMN: 3 mm [2-5], IF: 2 mm [1-3], CEMNIF: 3 mm [2-4]) were similar. Radiation exposure (EMN: 0; IF: 708 mGy.cm [599-1128]; CEMNIF: 51 mGy.cm [15-150]; p < 0.001) was significantly higher with IF than with CEMNIF and EMN. For targets with large angle access, procedure times (EMN: 345 s [259-457], CEMNIF: 425 s [340-473]; p = 0.01) and radiation exposure (EMN: 0, CEMIF: 159 mGy.cm [39-316]; p < 0.001) were significantly lower with EMN but with lower accuracy (EMN: 4 mm [4-6] and CEMNIF: 4 mm [3, 4]; p = 0.01). The operator's experience did not impact the tested parameters regardless of the technique.

Conclusion: In this phantom study, EMN was not limited to acute angle targets. Efficacy and accuracy of puncture for acute angle access targets with EMN, IF or CEMNIF were similar. CEMNIF is more accurate for large angle access targets at the cost of a slightly higher procedure time and radiation exposure.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00270-019-02356-wDOI Listing

Publication Analysis

Top Keywords

angle access
16
acute angle
12
electromagnetic navigation
8
image fusion
8
fusion fluoroscopy
8
phantom study
8
access targets
8
radiation exposure
8
emn
6
access
5

Similar Publications

Nutcracker syndrome (NCS) is an uncommon vascular condition caused by the compression of the left renal vein (LRN), which may result in venous hypertension and clinical symptoms, including hematuria, flank pain, and pelvic congestion. This report describes the case of a 30-year-old woman with recurrent painless macroscopic hematuria, ultimately diagnosed with NCS based on clinical and radiological findings. Computed tomography (CT) angiography revealed a reduced aortomesenteric angle and LRN compression without signs of severe venous hypertension or collateral circulation.

View Article and Find Full Text PDF

The choroid, a critical vascular layer beneath the retina, is essential for maintaining retinal function and monitoring chorioretinal disorders. Existing imaging methods, such as indocyanine green angiography (ICGA) and optical coherence tomography (OCT), face significant limitations, including contrast agent requirements, restricted field of view (FOV), and high costs, limiting accessibility. To address these challenges, we developed a nonmydriatic, contrast agent-free fundus camera utilizing transcranial near-infrared (NIR) illumination.

View Article and Find Full Text PDF

NMR spectroscopy presents boundless opportunities for understanding the structure, dynamics, and function for a broad range of scientific applications. Solid-state NMR (SSNMR), in particular, provides novel insights into biological and material systems that are not amenable to other approaches. However, a major bottleneck is the extent of user training and the difficulty of obtaining reproducible, high-quality experimental results, especially for the sophisticated multidimensional pulse sequences that are essential to provide site-resolved measurements in large biomolecules.

View Article and Find Full Text PDF

Introduction In idiopathic scoliosis surgery, studies have shown two attending surgeons have better curve correction, pain, and recovery time. There is conflicting evidence on operative time, blood loss, infection rate, and hospital length of stay. Limited literature examines the impact of surgeon experience on the dual approach.

View Article and Find Full Text PDF

Electrochemical Small-Angle X-ray Scattering for Potential-Dependent Structural Analysis of Redox Enzymes.

Langmuir

January 2025

Department of Pure and Applied Chemistry, Faculty of Science and Technology, Tokyo University of Science, 2641 Yamazaki, Noda, Chiba 278-8510, Japan.

Various methods exist for exploring different aspects of these mechanisms. However, techniques for investigating structural differences between the reduced and oxidized forms of an enzyme are limited. Here, we propose electrochemical small-angle X-ray scattering (EC-SAXS) as a novel method for potential-dependent structural analysis of redox enzymes and redox-active proteins.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!