kEDTA Sample Contamination: A Reappraisal.

J Appl Lab Med

Department of Blood Sciences, New Cross Hospital, Wolverhampton, UK.

Published: May 2019

Background: Potassium EDTA (kEDTA) contamination of serum samples is common, causing spurious hyperkalemia, hypozincemia, and hypocalcemia that if unrecognized may adversely affect patient care. Gross kEDTA contamination is easy to detect, but identification of spurious electrolytes due to small amounts of contamination requires measurement of serum EDTA. We validated an EDTA assay on the Abbott Architect and reassessed its value in identifying kEDTA contamination and in studying mechanisms for contamination.

Methods: Within- and between-batch imprecision, linearity, recovery, interference, and carryover were assessed. Serum supplemented with kEDTA plasma, to mimic sample contamination, was used to study its effect on potassium, calcium, zinc, magnesium, and alkaline phosphatase. Our current laboratory protocol for identification of kEDTA contamination, based on measurement of serum calcium, was compared to that of EDTA measurement.

Results: The EDTA assay displayed acceptable performance characteristics. Hemoglobin was a positive interferent. EDTA was detectable in serum contaminated with 1% (v:v) kEDTA plasma. An increase in serum potassium of 0.54 mmol/L (11.9%) was observed at a measured EDTA concentration of 0.19 mmol/L, equivalent to 3.2% (v:v) contamination. At this EDTA concentration reductions were also observed in zinc (71%), calcium (1%), alkaline phosphatase (ALP) (4%), and magnesium (2.4%). The serum EDTA assay detected contamination in 31/106 patient samples with hyperkalemia (potassium ≥6.0mmol/L), 20 of which were undetected by the current laboratory protocol.

Conclusions: The EDTA assay displayed acceptable performance, with the ability to reliably measure EDTA at low concentrations. Only a small amount of kEDTA causes significant spurious hyperkalemia and is only reliably detected with EDTA measurement.

Download full-text PDF

Source
http://dx.doi.org/10.1373/jalm.2018.027920DOI Listing

Publication Analysis

Top Keywords

kedta contamination
16
edta assay
16
edta
12
contamination
9
kedta
8
sample contamination
8
spurious hyperkalemia
8
measurement serum
8
serum edta
8
kedta plasma
8

Similar Publications

Introduction: A specific sequence is recommended for filling blood tubes during blood collection to prevent erroneous test results due to carryover of additives. However, requirement of this procedure is still debatable. This study was aimed to investigate the potassium ethylenediaminetetraacetic acid (K-EDTA) contamination in blood samples taken after a tube containing the additive during routine workflow.

View Article and Find Full Text PDF

kEDTA Sample Contamination: A Reappraisal.

J Appl Lab Med

May 2019

Department of Blood Sciences, New Cross Hospital, Wolverhampton, UK.

Background: Potassium EDTA (kEDTA) contamination of serum samples is common, causing spurious hyperkalemia, hypozincemia, and hypocalcemia that if unrecognized may adversely affect patient care. Gross kEDTA contamination is easy to detect, but identification of spurious electrolytes due to small amounts of contamination requires measurement of serum EDTA. We validated an EDTA assay on the Abbott Architect and reassessed its value in identifying kEDTA contamination and in studying mechanisms for contamination.

View Article and Find Full Text PDF

Aim: To investigate whether incorrect order of draw of blood samples during phlebotomy causes in vitro potassium ethylenediaminetetraacetic acid EDTA (kEDTA) contamination of blood samples.

Methods: Serum kEDTA, potassium, calcium, magnesium, alkaline phosphatase, zinc and iron concentrations were measured in blood samples drawn before and after collecting blood into kEDTA containing sample tubes by an experienced phlebotomist using the Sarstedt Safety Monovette system.

Results: EDTA was undetectable in all samples.

View Article and Find Full Text PDF

Background: A multi-centre observational study investigating the prevalence of spurious hyperkalaemia due to potassium ethylenediaminetetraacetic acid (kEDTA) contamination.

Methods: Serum EDTA was measured in anonymised serum samples with a serum potassium > 6.0 mmol/L collected over a one month period in five different hospital laboratories.

View Article and Find Full Text PDF

Characterizing the effects of shaking intensity on the kinetics of metallic iron dissolution in EDTA.

J Hazard Mater

October 2009

Angewandte Geologie, Universität Göttingen, Goldschmidtstrasse 3, D-37077 Göttingen, Germany.

Despite two decades of intensive laboratory investigations, several aspects of contaminant removal from aqueous solutions by elemental iron materials (e.g., in Fe(0)/H2O systems) are not really understood.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!