Background: Endotoxin contamination of dialysate has serious adverse effects on patients undergoing hemodialysis. Therefore, endotoxin activity in dialysate is closely monitored. Limulus amebocyte lysate (LAL) has been used as a reagent to measure endotoxin activity. Here, we investigated the efficacy of an automatic LAL kinetic turbidimetric test (Toxinometer ET-mini) for screening endotoxin activity in dialysate.

Methods: In total, endotoxin activity was measured in 110 dialysate samples obtained from several sites within hemodialysis circuits between June 2012 and March 2018. The results were compared with those from a conventional chromogenic substrate LAL test conducted by a clinical examination laboratory.

Results: Both the automatic LAL test and the chromogenic substrate LAL test had a minimum detection level of 0.001 endotoxin units (EU)/mL. Endotoxin activity levels measured via the automatic LAL test showed a strongly positive correlation (concordance correlation coefficient: 0.9933; 95% CI: 0.9902-0.9954) and good agreement (mean difference: 0.00±0.01 EU/mL) with those obtained using the chromogenic substrate LAL test.

Conclusion: The results suggest that the automatic LAL test may be useful for endotoxin activity screening in hemodialysis facilities.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC6789175PMC
http://dx.doi.org/10.2147/MDER.S225246DOI Listing

Publication Analysis

Top Keywords

endotoxin activity
24
lal test
20
automatic lal
16
chromogenic substrate
12
substrate lal
12
endotoxin
9
limulus amebocyte
8
amebocyte lysate
8
kinetic turbidimetric
8
turbidimetric test
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!