Purpose: The purpose of this study was to investigate the feasibility of two-dimensional (2D) dose distribution deconvolution using convolutional neural networks (CNNs) instead of an analytical approach for an in-house scintillation detector that has a detector-interface artifact in the penumbra region.

Methods: Datasets of 2D dose distributions were acquired from a medical linear accelerator of Novalis Tx. The datasets comprise two different sizes of square radiation fields and 13 clinical intensity-modulated radiation treatment (IMRT) plans. These datasets were divided into two datasets (training and test) to train and validate the developed network, called PenumbraNet, which is a shallow linear CNN. The PenumbraNet was trained to transform the measured dose distribution [M(x, y)] to calculated distribution [D(x, y)] by the treatment planning system. After training of the PenumbraNet was completed, the performance was evaluated using test data, which were 10 × 10 cm open field and ten clinical IMRT cases. The corrected dose distribution [C(x, y)] was evaluated against D(x, y) with 2%/2 mm and 3%/3 mm criteria of the gamma index for each field. The M(x, y) and deconvolved dose distribution with the analytically obtained kernel using Wiener filtering [A(x, y)] were also evaluated for comparison. In addition, we compared the performance of the shallow depth of linear PenumbraNet with that of nonlinear PenumbraNet and a deep nonlinear PenumbraNet within the same training epoch.

Results: The mean gamma passing rates were 84.77% and 95.81% with 3%/3 mm gamma criteria for A(x, y) and C(x, y) of the PenumbraNet, respectively. The mean gamma pass rates of nonlinear PenumbraNet and the deep depth of nonlinear PenumbraNet were 96.62%, 93.42% with 3%/3 mm gamma criteria, respectively.

Conclusions: We demonstrated the feasibility of the PenumbraNets for 2D dose distribution deconvolution. The nonlinear PenumbraNet which has the best performance improved the gamma passing rate by 11.85% from the M(x, y) at 3%/3 mm gamma criteria.

Download full-text PDF

Source
http://dx.doi.org/10.1002/mp.13869DOI Listing

Publication Analysis

Top Keywords

dose distribution
24
nonlinear penumbranet
20
distribution deconvolution
12
3%/3 mm gamma
12
gamma criteria
12
penumbranet
10
feasibility two-dimensional
8
two-dimensional dose
8
neural networks
8
penumbranet deep
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!