Introduction: Positive pressure ventilation (PPV) with T-Piece and self-inflating bag (SIB) during neonatal resuscitation after birth is associated with variability in ventilation. The use of a ventilator with respiratory function monitoring (RFM) for PPV, however, has not been evaluated.
Objective: To determine if ventilator + RFM can reduce ventilation variability compared to T-Piece and SIB in a preterm manikin at different combinations of target tidal volume (VT) and lung compliance (CL).
Methods: Twenty clinicians provided PPV via mask and endotracheal tube (ETT) using SIB, T-Piece, T-Piece + RFM and Ventilator + RFM to a manikin with adjustable lung CL. Three combinations of CL and target VT: Low CL-Low VT, Low CL-High VT and High CL-Low VT were used in a random order.
Results: The use of ventilator + RFM for PPV via ETT during High CL-Low VT period reduced the proportion of breaths with expiratory VT above target when compared to the other 3 devices (56 ± 35%, 85 ± 20%, 90 ± 25%, 92 ± 12% for ventilator + RFM, T-Piece + RFM, T-Piece, SIB, respectively; p < 0.05). During PPV via both mask and ETT, ventilator + RFM maintained the set Ti and rate, whereas SIB and T-Piece use resulted in higher rates, and T-Piece in higher proportion of breaths with prolonged Ti. During PPV via mask, ventilator + RFM reduced gas leakage compared to other devices.
Conclusion: In this simulation study, use of a mechanical ventilator with RFM led to an overall improvement in volume targeting at different settings of CL and reduced the gas leak during mask ventilation. The efficacy and safety of using this strategy to neonatal resuscitation in the delivery room needs to be evaluated.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1159/000503257 | DOI Listing |
J Perinat Med
November 2024
Division of Neonatology, Center for Maternal-Fetal, Neonatal and Reproductive Medicine, 13611 National Center for Child Health and Development, Setagaya-ku, Tokyo, Japan.
Objectives: To investigate the effectiveness of placing an indicator light indicating inadequate ventilation near the face mask during positive-pressure ventilation with respiratory function monitors (RFMs) in neonatal resuscitation. The study is a three-group, randomised, controlled, crossover simulation trial.
Methods: Paediatrics residents, neonatology fellows, and attending neonatologists at a single neonatal intensive care unit.
Resuscitation
October 2024
Division of Neonatology, Pediatric Intensive Care and Neuropediatrics, Department of Pediatrics, Comprehensive Center for Pediatrics, Medical University of Vienna, Vienna, Austria. Electronic address:
Objective: The aim of this study was the evaluation of the impact of a respiratory function monitor (RFM, Neo100, Monivent AB, Gothenburg, Sweden) on the quality of ventilation in neonates.
Methods: This single-center two-phase intervention study was conducted at the Neonatal Intensive Care Unit and the delivery room of the Medical University of Vienna. Patients with clinical need for positive pressure ventilation were included in either of two consecutive study phases: (i) patients were ventilated with a hidden RFM (control) or (ii) visible RFM (intervention) during manual positive pressure ventilations.
Resusc Plus
June 2024
Department of Neonatology, National Maternity Hospital, Dublin, Ireland.
Objective: One in twenty newborns require resuscitation with positive pressure ventilation (PPV) at birth. Newborn face mask ventilation is often poorly performed. To address this, the potential role of respiratory function monitors (RFM) in newborn resuscitation training has been highlighted.
View Article and Find Full Text PDFArch Dis Child Fetal Neonatal Ed
August 2024
Neonatal Intensive Care Unit, Westmead Hospital, Sydney, New South Wales, Australia.
Background: Respiratory function monitors (RFMs) have been used extensively in manikin and infant studies yet have not become the standard of training. We report the outcomes of a new portable, lightweight RFM, the Juno, designed to show mask leak and deflation tidal volume to assist in positive pressure ventilation (PPV) competency training using manikins.
Methods: Two leak-free manikins (preterm and term) were used.
Arch Dis Child Fetal Neonatal Ed
August 2024
The National Maternity Hospital, Dublin, Ireland.
Objective: The International Liaison Committee on Resuscitation has recommended improvements in training for neonatal resuscitation, highlighting the potential role of respiratory function monitors (RFMs). Our objective was to determine whether a manikin-based, standardised face mask ventilation training intervention using an RFM with a simple visual display reduced face mask leak.
Design: Multicentre, before and after study.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!