Purpose: QuantiFERON-TB Gold Plus (QFT-Plus) is a new generation of QuantiFERON assay that differs from QuantiFERON-TB Gold In-Tube test (QFT-GIT). The aim of this study was to compare the performance of the new FDA-approved QFT-Plus interferon (IFN)-γ release assays (IGRAs) with the QFT-GIT version of this assay.

Material And Methods: We searched all studies published in English in electronic databases, including PubMed, Scopus, and Web of Science.

Results: The positive proportion of positive results by QFT-Plus was higher than QFT-GIT in cured tuberculosis (TB) cases (82% vs. 73%). The two tests showed a substantial agreement and the majority of the latent tuberculosis infection (LTBI) subjects responded concomitantly to both QFT-Plus and QFT-GIT. However, QFT-Plus showed a stronger association with surrogate measures of TB suspects than QFT-GIT. The QFT-Plus test demonstrated a higher sensitivity than QFT-GIT in the older adults. The sensitivity, specificity, LR+, LR- and DOR overall were 94% (95% CI 89-97), 96% (95% CI 94-98), 24.4 (95% CI 15-39), 0.05 (95% CI 0.03-0.11) and 414 (95% CI 251-685), respectively. The area under summary ROC curve was 0.99 (95% CI 0.97-0.99).

Conclusion: QFT-Plus performs equivalently to the QFT-GIT for detection of patients at risk for LTBI; however, QFT-Plus test had higher sensitivity than the QFT-GIT test, with similar specificity among the older participants. Higher IFN-γ release in TB2 compared to TB1 might be due to recent LTBI.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.advms.2019.09.001DOI Listing

Publication Analysis

Top Keywords

quantiferon-tb gold
16
gold in-tube
8
release assays
8
qft-plus
8
qft-git
8
ifn-γ release
8
qft-git qft-plus
8
qft-plus test
8
higher sensitivity
8
sensitivity qft-git
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!