Impact of matrix effects and ionization efficiency in non-quantitative untargeted metabolomics.

Metabolomics

Department of Pathology, Immunology and Laboratory Medicine, University of Florida, Gainesville, FL, 32610, USA.

Published: October 2019

Introduction: LC-MS-based untargeted metabolomics has become increasingly popular due to the vast amount of information gained in a single analysis. Many studies utilize metabolomics to profile metabolic changes in various representative biofluids, tissues, or other sample types. Most analyses are performed measuring changes in the metabolic pool of a single biological matrix due to an altered phenotype, such as disease versus normal. Measurements in such experiments are typically highly reproducible with little variation due to analytical and technological advancements in mass spectrometry. With the expanded application of metabolomics into various non-analytical scientific disciplines, the emergence of studies comparing the signal intensities of specific analytes across different biological matrices (e.g. plasma vs. urine) is becoming more common, but the matrix effect between sample types is often neglected. Additionally, the practice of comparing the signal intensities of different analytes and correlating to relative abundance is also increasingly prevalent, but the response ratio between analytes due to differences in ionization efficiency is not always accounted for in data analysis. This report serves to communicate and raise awareness of these two well-recognized issues to prevent improper data interpretation in the field of metabolomics.

Objectives: We demonstrate the impact of matrix effects and ionization efficiency with labeled analytical standards in human plasma, serum, and urine and describe how the direct comparison of non-quantitative signal intensities between biofluids, as well as between different analytes in the same biofluid, in untargeted metabolomics is inaccurate without proper response corrections.

Methods: Human plasma, serum, and urine (n = 4 technical replicates per biofluid) were spiked with a panel of labeled internal standards all at identical concentrations, simultaneously extracted, and analyzed by UHPLC-HRMS. Signal intensities were compared for demonstration of the impact of matrix effects in untargeted metabolomics. A neat mixture of two co-eluting, structurally-similar labeled standards at the same concentration was also analyzed to demonstrate the effect of ionization efficiency on signal intensity.

Results: Despite being spiked at identical concentrations, labeled standards we examined in this study showed significant differences in their signal intensities between biofluids, as well as from each other in the same biofluid, due to matrix effects. Co-eluting standards were also found to yield significantly different signal intensities at identical concentrations due to differences in ionization efficiency.

Conclusions: Due to the presence of matrix effects in untargeted, non-quantitative metabolomics, the signal intensity of any single analyte cannot be directly compared to the signal intensity of that same analyte (or any other analyte) between any two different matrices. Due to differences in ionization efficiency, the signal intensity of any single analyte cannot be directly compared to the signal intensity of any other analyte, even in the same matrix.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s11306-019-1597-zDOI Listing

Publication Analysis

Top Keywords

signal intensities
24
matrix effects
20
ionization efficiency
20
untargeted metabolomics
16
signal intensity
16
impact matrix
12
differences ionization
12
identical concentrations
12
signal
11
effects ionization
8

Similar Publications

Visualizing enterohepatic circulation in vivo by sensitive F MRI with a fluorinated ferrous chelate-based small molecule probe.

Biomaterials

January 2025

The Key Laboratory for Chemical Biology of Fujian Province, The MOE Key Laboratory of Spectrochemical Analysis & Instrumentation, and Department of Chemical Biology, College of Chemistry and Chemical Engineering, Xiamen University, Xiamen, 361005, China; Fujian Provincial Key Laboratory of Chronic Liver Disease and Hepatocellular Carcinoma, Xiamen Key Laboratory of Translational Medical of Digestive System Tumor, Zhongshan Hospital, Xiamen University, Xiamen 361004, China. Electronic address:

Enterohepatic circulation (EHC) is a critical biological process for the normal regulation of many endogenous biomolecules and the increased retention of various exogenous substances. The status of EHC is closely related to the ordinary functioning of several digestive organs. However, it remains a challenge to achieve in vivo real-time visualization of this process.

View Article and Find Full Text PDF

Gadopiclenol Enables Reduced Gadolinium Dose While Maintaining Quality of Pulmonary Arterial Enhancement for Pulmonary MRA: An Opportunity for Improved Safety and Sustainability.

Invest Radiol

January 2025

From the Departments of Radiology (J.F.H., S.Y.C., J.-P.G., J.S., P.N., S.B.R., T.M.G.), Biomedical Engineering (S.B.R., T.M.G.), Medical Physics (S.Y.C., S.B.R., T.M.G.), Medicine (S.B.R.), and Emergency Medicine (S.B.R.), University of Wisconsin-Madison, WI; and Department of Diagnostic and Interventional Radiology (J.F.H., J.-P.G.), University Hospital Würzburg, Würzburg, Germany.

Rationale And Objectives: Pulmonary magnetic resonance angiography (MRA) is an imaging method with proven utility for the exclusion of pulmonary embolism and avoids the need for ionizing radiation and iodinated contrast agents. High-relaxivity gadolinium-based contrast agents (GBCAs), such as gadopiclenol, can be used to reduce the required gadolinium dose for pulmonary MRA. The aim of this study was to compare the contrast enhancement performance of gadopiclenol with an established gadobenate dimeglumine-enhanced pulmonary MRA protocol.

View Article and Find Full Text PDF

Sleep need accumulates during waking and dissipates during sleep to maintain sleep homeostasis (process S). Besides the regulation of daily (baseline) sleep amount, homeostatic sleep regulation commonly refers to the universal phenomenon that sleep deprivation (SD) causes an increase of sleep need, hence, the amount and intensity of subsequent recovery sleep. The central regulators and signaling pathways that govern the baseline and homeostatic sleep regulations in mammals remain unclear.

View Article and Find Full Text PDF

Background: While the effects of sleep deprivation on cognitive function are well-documented, its impact on high-intensity endurance performance and underlying neural mechanisms remains underexplored, especially in the context of search and rescue operations where both physical and mental performance are essential. This study examines the neurophysiological basis of sleep deprivation on high-intensity endurance using electroencephalography (EEG). In this crossover study, twenty firefighters were subjected to both sleep deprivation (SD) and normal sleep conditions, with each participant performing endurance treadmill exercise the following morning after each condition.

View Article and Find Full Text PDF

We present a rare case of a patient with co-occurring exercise-induced acute kidney injury (AKI) and rhabdomyolysis. A 67-year-old man was referred to our department with AKI. Five days before referral, the patient had sudden-onset loin pain while banging and kicking on a door in a holding cell at a police station.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!