Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objectives: This study assessed the long-term durability of nominally deployed transcatheter heart valves (THV) to 1 billion cycles (equivalent to 25 years) and non-nominal (overexpansion, underexpansion, and elliptical) THV deployments to 200 million cycles (equivalent to 5 years) with accelerated wear testing.
Background: The long-term durability of THVs is currently unknown. As transcatheter aortic valve replacement expands to lower-risk patients, durability will be of increasing importance.
Methods: SAPIEN 3 THVs, sized 20, 23, 26, and 29 mm were assessed. Nominally deployed THVs underwent hydrodynamic performance and mechanical durability as assessed with accelerated wear testing to 1 billion cycles. Magna Ease surgical valves were used as comparators. Durability of non-nominal THV deployments was tested to 200 million cycles. Valves were tested to International Standards Organization 5840:2013 standard.
Results: THV durability was excellent for both the nominal and non-nominal THV deployments to 1 billion and 200 million cycles, respectively. At 1 billion cycles the regurgitant fraction for the 20-, 23-, 26-, and 29-mm SAPIEN 3 was 0.92 ± 0.47%, 1.29 ± 0.04%, 1.73 ± 0.46%, and 2.47 ± 0.15%, respectively. There was also excellent durability in the comparator Magna Ease valves. The regurgitant fraction of non-nominal overexpanded (20 mm, 4.36 ± 0.53; 23 mm, 7.68 ± 1.39; 26 mm, 6.80 ± 1.17; 29 mm, 9.00 ± 0.37), underexpanded (20 mm, 3.06 ± 0.28; 23 mm, 4.46 ± 0.45; 26 mm, 7.72 ± 0.48; 29 mm, 8.65 ± 2.01), and elliptical (20 mm, 3.30 ± 0.38; 23 mm, 6.13 ± 0.94; 26 mm, 6.77 ± 1.22; 29 mm, 8.72 ± 0.24) THVs were excellent at 200 million cycles.
Conclusions: Nominal SAPIEN 3 THVs demonstrated excellent durability, to an equivalent of 25-years wear. THV durability was similar to the comparator surgical valves tested. Non-nominal (overexpansion, underexpansion, and elliptical) THV deployments also had excellent durability to an equivalent of 5 years wear.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jcin.2019.07.049 | DOI Listing |