Reproducibility of Natural Shear Wave Elastography Measurements.

Ultrasound Med Biol

Cardiology, Thorax Center, Erasmus MC, Rotterdam, The Netherlands; Acoustical Wavefield Imaging, ImPhys, Delft University of Technology, The Netherlands.

Published: December 2019

For the quantification of myocardial function, myocardial stiffness can potentially be measured non-invasively using shear wave elastography. Clinical diagnosis requires high precision. In 10 healthy volunteers, we studied the reproducibility of the measurement of propagation speeds of shear waves induced by aortic and mitral valve closure (AVC, MVC). Inter-scan was slightly higher but in similar ranges as intra-scan variability (AVC: 0.67 m/s (interquartile range [IQR]: 0.40-0.86 m/s) versus 0.38 m/s (IQR: 0.26-0.68 m/s), MVC: 0.61 m/s (IQR: 0.26-0.94 m/s) versus 0.26 m/s (IQR: 0.15-0.46 m/s)). For AVC, the propagation speeds obtained on different day were not statistically different (p = 0.13). We observed different propagation speeds between 2 systems (AVC: 3.23-4.25 m/s [Zonare ZS3] versus 1.82-4.76 m/s [Philips iE33]), p = 0.04). No statistical difference was observed between observers (AVC: p = 0.35). Our results suggest that measurement inaccuracies dominate the variabilities measured among healthy volunteers. Therefore, measurement precision can be improved by averaging over multiple heartbeats.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ultrasmedbio.2019.09.002DOI Listing

Publication Analysis

Top Keywords

propagation speeds
12
m/s iqr
12
m/s
10
shear wave
8
wave elastography
8
healthy volunteers
8
m/s versus
8
avc
5
reproducibility natural
4
natural shear
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!